Assessing the validity of pressure-measuring insoles in quantifying gait variables
Journal of Rehabilitation and Assistive Technologies Engineering · 2018-01 · 5
DOI: 10.1177/2055668317752088 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Rehabilitation and Assistive Technologies Engineering, online erschienen am 12.01.2018. Verfasst von Jessica DeBerardinis und 3 weiteren Autor:innen. Beteiligte Einrichtung: 1Department of Mechanical Engineering, University of Nevada Las Vegas, Las Vegas, USA. Die Arbeit wurde bisher 27-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: University of Nevada, Las Vegas.
Abstract (Originalsprache)
Introduction
Pressure-measuring insoles can provide a portable alternative to existing gait analysis tools. However, there is disagreement among researchers on their accuracy and the appropriate calibration methods. The purposes of this study were to (1) determine the validity of pressure-measuring insoles for calculating stance time and support-phase impulse during walking using two calibration procedures, and (2) examine the effect of insole size on the results.
Methods
Data were collected from 39 participants (23.5 ± 3.24 yrs, 66.7 ± 17.5 kg, 1.64 ± 0.09 m), each wearing appropriately sized insoles as they walked over two consecutive force platforms. Two calibration methods were evaluated: (1) manufacturer’s recommendation, and (2) a participant weight-based approach. Qualitative and quantitative evaluations were conducted.
Results
The results indicated that the insoles measured longer stance times than the force platform (differences are less than 10%). Both calibration methods resulted in inaccurate impulse values (differences are 30 and 50% for the two calibration methods, respectively). The results showed that when using the first calibration method, impulse values depended on insole size. The second calibration consistently underestimated the impulse.
Conclusions
It was concluded that while the insoles provide acceptable qualitative representation of the gait, the two studied calibration methods may lead to a misleading quantitative assessment.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- 1Department of Mechanical Engineering, University of Nevada Las Vegas, Las Vegas, USA.
- 2Department of Kinesiology and Nutrition Sciences, University of Nevada Las Vegas, Las Vegas, USA.
ORCID-Profile: Daniel E Lidstone
Förderung
- University of Nevada, Las Vegas
Zitieren
APA
DeBerardinis, J., Dufek, J. S., Trabia, M. B., & Lidstone, D. E. (2018). Assessing the validity of pressure-measuring insoles in quantifying gait variables. Journal of Rehabilitation and Assistive Technologies Engineering, 5. https://doi.org/10.1177/2055668317752088
Vancouver
DeBerardinis J, Dufek JS, Trabia MB, Lidstone DE. Assessing the validity of pressure-measuring insoles in quantifying gait variables. Journal of Rehabilitation and Assistive Technologies Engineering. 2018;5. doi:10.1177/2055668317752088
BibTeX
@article{DeBerardinis2018,
author = {DeBerardinis, Jessica and Dufek, Janet S and Trabia, Mohamed B and Lidstone, Daniel E},
title = {Assessing the validity of pressure-measuring insoles in quantifying gait variables},
journal = {Journal of Rehabilitation and Assistive Technologies Engineering},
year = {2018},
volume = {5},
doi = {10.1177/2055668317752088}
}
EUVIMED