Validation of wearable vital signs monitoring: A comparison with conventional bedside patient monitors

Weiyi Jiang iD, Wenzhao Zhang, Haoxuan Li iD, Jean Ngoie, Wenda Li iD, Zhihong Huang

DIGITAL HEALTH · 2025-05 · 11

DOI: 10.1177/20552076251377934 ↗

Kurz gefasst

Fachartikel in der Zeitschrift DIGITAL HEALTH, online erschienen am 03.11.2025. Verfasst von Weiyi Jiang und 5 weiteren Autor:innen. Beteiligte Einrichtung: School of Physics, Engineering and Technology, University of York, York, UK. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

This study aimed to assess how accurately mobile wearable devices could be used to monitor vital signs continuously in clinical settings by comparing their measurements with those from traditional bedside monitors.

Methods

Data collected from Mindray's mWear wearable device were compared against measurements from the BeneVision N15 traditional bedside monitoring system. A total of 208 paired datasets, including blood pressure, oxygen saturation, heart rate, and respiratory rate, were collected from 16 healthy volunteers in a clinical setting. Bland-Altman analysis was applied to assess agreement between the two devices.

Results

The analysis showed that 94.2% of the data variance points for oxygen saturation, diastolic blood pressure, and pulse rate fell within the limits of agreement. For systolic blood pressure, 92.3% of the data variance points were within limits, while 94.7% of the heart rate and respiratory rate data points were also within agreement limits.

Conclusion

There was a strong agreement between the wearable mWear device and the traditional bedside patient monitoring system. This study supports the accuracy and reliability of wearable devices for continuous vital signs monitoring. These results encourage the wider use and ongoing improvement of wearable technology in clinical practice.

Schlagwörter

Stichwörter der Autor:innen

Bland-Altman analysisVital signsdigital healthnewborn healthpatient monitoringwearableswireless

Autor:innen und Institutionen

  • School of Physics, Engineering and Technology, University of York, York, UK.
  • Institute of Signals, Sensors and Systems, Heriot-Watt University, Edinburgh, UK.
  • CHEO Research Institute, Ottawa, Ontario, Canada.

ORCID-Profile: Weiyi Jiang, Haoxuan Li, Wenda Li

Zitieren

APA

Jiang, W., Zhang, W., Li, H., Ngoie, J., Li, W., & Huang, Z. (2025). Validation of wearable vital signs monitoring: A comparison with conventional bedside patient monitors. DIGITAL HEALTH, 11. https://doi.org/10.1177/20552076251377934

Vancouver

Jiang W, Zhang W, Li H, Ngoie J, Li W, Huang Z. Validation of wearable vital signs monitoring: A comparison with conventional bedside patient monitors. DIGITAL HEALTH. 2025;11. doi:10.1177/20552076251377934

BibTeX
@article{Jiang2025,
  author = {Jiang, Weiyi and Zhang, Wenzhao and Li, Haoxuan and Ngoie, Jean and Li, Wenda and Huang, Zhihong},
  title = {Validation of wearable vital signs monitoring: A comparison with conventional bedside patient monitors},
  journal = {DIGITAL HEALTH},
  year = {2025},
  volume = {11},
  doi = {10.1177/20552076251377934}
}
Kostenlos starten →