Influence of tube voltage on image quality in non-contrast photon-counting computed tomography of the head: Comparison of 120 kVp and 140 kVp
The Neuroradiology Journal · 2025
DOI: 10.1177/19714009251339079 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift The Neuroradiology Journal, 2025. Verfasst von Denise Schoenbeck und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Neuroradiology and Nuclear Medicine, Johannes Wesling University Hospital, Germany. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Purpose
Non-contrast cerebral computed tomography (NCCT) is one of the most frequently performed CT examinations. Photon-counting CT (PCCT) offers advantages in terms of noise reduction, higher spatial resolution, and inherent spectral information. PCCT available today allows NCCT to be performed with tube voltage of 120 or 140 kVp. This study evaluates the impact of tube voltage on image quality at an equivalent dose.
Methods
76 patients with an NCCT with 120 kVp, 76 with 140 kVp, and 56 patients with slightly different effective tube current per group were included. Signal, noise, signal-to-noise ratio, gray–white contrast, and contrast-to-noise ratio were determined using several regions of interest for different virtual monoenergetic image (VMI) levels and compared between dose-equivalent groups. An image quality rating of the clinically used virtual monoenergetic images (VMIs) 65 keV was performed.
Results
The VMI 65 keV images at 120 kVp exhibited reduced noise, improved gray–white contrast, and improved contrast-to-noise ratio compared to 140 kVp ( p < .001). The density differences between cortical gray matter at different distances from calvaria were also lower with 120 kVp ( p < .001). The rating of image quality showed no difference between 120 kVp and 140 kVp.
Conclusions
Currently, NCCT with a tube voltage of 120 kVp versus 140 kVp seems to achieve better image quality. However, further studies are required to evaluate possible advantages of 140 kVp, for example artifact reduction in the case of dense foreign materials or enhanced spectral possibilities, and regarding imaging of special intracranial pathologies.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, Neuroradiology and Nuclear Medicine, Johannes Wesling University Hospital, Germany.
- Ruhr University Bochum, Germany.
- Department of Radiology and Neuroradiology, University Hospital of Cologne, Germany.
ORCID-Profile: Denise Schoenbeck
Zitieren
APA
Schoenbeck, D., Kroeger, J. R., Woeltjen, M. M., Niehoff, J. H., Moenninghoff, C., Goertz, L., Borggrefe, J., & Michael, A. E. (2025). Influence of tube voltage on image quality in non-contrast photon-counting computed tomography of the head: Comparison of 120 kVp and 140 kVp. The Neuroradiology Journal. https://doi.org/10.1177/19714009251339079
Vancouver
Schoenbeck D, Kroeger JR, Woeltjen MM, Niehoff JH, Moenninghoff C, Goertz L, et al. Influence of tube voltage on image quality in non-contrast photon-counting computed tomography of the head: Comparison of 120 kVp and 140 kVp. The Neuroradiology Journal. 2025. doi:10.1177/19714009251339079
BibTeX
@article{Schoenbeck2025,
author = {Schoenbeck, Denise and Kroeger, Jan Robert and Woeltjen, Matthias Michael and Niehoff, Julius Henning and Moenninghoff, Christoph and Goertz, Lukas and Borggrefe, Jan and Michael, Arwed Elias},
title = {Influence of tube voltage on image quality in non-contrast photon-counting computed tomography of the head: Comparison of 120 kVp and 140 kVp},
journal = {The Neuroradiology Journal},
year = {2025},
doi = {10.1177/19714009251339079}
}
EUVIMED