Fachartikel in der Zeitschrift Journal of Diabetes Science and Technology, online erschienen am 19.07.2026. Verfasst von Fengyun Wang und 4 weiteren Autor:innen. Beteiligte Einrichtung: Children's Hospital of Soochow University, Suzhou, China. Die Arbeit nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Sinocare Inc..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

A proprietary continuous glucose monitoring (CGM) system that uses third-generation sensor technology has previously been shown to exhibit stable, single-digit mean absolute relative difference (MARD) and 15-day wear life in adults with diabetes.

Method

This was a prospective, single-arm study of pediatric participants aged 2 to 17 years that were enrolled at 3 clinical centers. Participants wore 2 sensors, one on each side of the abdomen, and were randomized to assessment of device performance on days 1 and 2, days 7 to 9 or days 15 and 16. Device performance across a range of metrics was assessed by comparison with venous blood glucose values obtained using a laboratory reference device. Per-protocol results are reported for the sensor with the inferior MARD.

Results

The per-protocol set comprised 75 participants, of whom 16 were aged 2 to 5 years. The overall MARD was 8.89%, and MARDs at days 1 and 2, 7 to 9 and 15 and 16 were 8.65%, 8.70% and 9.30%, respectively. DTS error grid analyses showed that 100.0% of data pairs fell in clinically acceptable zones A+B. True alarm rates for hypoglycemia and hyperglycemia were high, at 98.6% and 98.8%, respectively. Mean sensor wear life was 14.4 days, and participant/guardian assessments of sensor usability revealed high satisfaction with respect to system assembly, sensor insertion and overall comfort. No device-related or skin-related adverse events were reported.

Conclusions

In a pediatric population, the novel CGM system demonstrated accurate and stable performance across its 15-day wear life and showed high usability.

Schlagwörter

Stichwörter der Autor:innen

continuous glucose monitormean absolute relative difference (MARD)pediatricssensor accuracywear life

Autor:innen und Institutionen

  • Children's Hospital of Soochow University, Suzhou, China.
  • Flacke Consulting GmbH, Mainz, Germany.
  • Sinocare Inc., Changsha, China.
  • Women and Children's Hospital of Ningbo University, China.
  • Hunan Children's Hospital, Changsha, China.

ORCID-Profile: Frank Flacke, Haiying Wu

Förderung

  • Sinocare Inc.

Zitieren

APA

Wang, F., Flacke, F., Ding, S., Oyang, W., & Wu, H. (2026). Performance, Safety and Usability of a Third-Generation Continuous Glucose Monitoring System in a Pediatric Population: Results From a Prospective, Multicenter Study. Journal of Diabetes Science and Technology. https://doi.org/10.1177/19322968261463536

Vancouver

Wang F, Flacke F, Ding S, Oyang W, Wu H. Performance, Safety and Usability of a Third-Generation Continuous Glucose Monitoring System in a Pediatric Population: Results From a Prospective, Multicenter Study. Journal of Diabetes Science and Technology. 2026. doi:10.1177/19322968261463536

BibTeX
@article{Wang2026,
  author = {Wang, Fengyun and Flacke, Frank and Ding, Shuxia and Oyang, Wenxian and Wu, Haiying},
  title = {Performance, Safety and Usability of a Third-Generation Continuous Glucose Monitoring System in a Pediatric Population: Results From a Prospective, Multicenter Study},
  journal = {Journal of Diabetes Science and Technology},
  year = {2026},
  doi = {10.1177/19322968261463536}
}
Kostenlos starten →