A Prospective Multicenter Assessment of the Accuracy and Safety of the Yuwell CT3 Real‐Time Continuous Glucose Monitoring System in Patients With Diabetes Over 14 Days
Journal of Diabetes · 2026-07 · 18 (7)
DOI: 10.1111/1753-0407.70246 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Journal of Diabetes, online erschienen am 14.07.2026. Verfasst von Lingli Zhou und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Endocrinology and Metabolism, Peking University People's Hospital, Beijing, China. Hauptthemen laut MEDLINE: Blood Glucose, Blood Glucose Self-Monitoring, Continuous Glucose Monitoring, Diabetes Mellitus. Die Arbeit nennt 13 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Science and Technology Major Project.
Themen: Diabetes mellitus
Abstract (Originalsprache)
Highlights This 14‐day multicenter study assessed CT3 CGM accuracy and safety in 71 adults with diabetes. Key results: overall MARD 9.1% (median 7.1%), 82.4%–98.2% met %15/15–%30/30 criteria, 99.7% values in CEG zones A/B. CT3 showed strong performance, suitable for reliable 14‐day wear.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Endocrinology and Metabolism, Peking University People's Hospital, Beijing, China.
- Department of Endocrinology, Beijing Pinggu Hospital, Beijing, China.
- Department of Endocrinology, Army General Hospital of Chinese PLA, Beijing, China.
- Department of Endocrinology, The Affiliated Hospital of Xuzhou Medical University, Xuzhou, Jiangsu, China.
- Department of Endocrinology, The Second Affiliated Hospital of Nanjing Medical University, Nanjing, China.
ORCID-Profile: Linong Ji
Förderung
- National Science and Technology Major Project
Zitieren
APA
Zhou, L., Li, Y., Zhang, X., Ling, H., Lu, Y., & Ji, L. (2026). A Prospective Multicenter Assessment of the Accuracy and Safety of the Yuwell CT3 Real‐Time Continuous Glucose Monitoring System in Patients With Diabetes Over 14 Days. Journal of Diabetes, 18(7). https://doi.org/10.1111/1753-0407.70246
Vancouver
Zhou L, Li Y, Zhang X, Ling H, Lu Y, Ji L. A Prospective Multicenter Assessment of the Accuracy and Safety of the Yuwell CT3 Real‐Time Continuous Glucose Monitoring System in Patients With Diabetes Over 14 Days. Journal of Diabetes. 2026;18(7). doi:10.1111/1753-0407.70246
BibTeX
@article{Zhou2026,
author = {Zhou, Lingli and Li, Yufeng and Zhang, Xingguang and Ling, Hongwei and Lu, Yibing and Ji, Linong},
title = {A Prospective Multicenter Assessment of the Accuracy and Safety of the Yuwell CT3 Real‐Time Continuous Glucose Monitoring System in Patients With Diabetes Over 14 Days},
journal = {Journal of Diabetes},
year = {2026},
volume = {18},
number = {7},
doi = {10.1111/1753-0407.70246}
}
EUVIMED