Multicenter Evaluation of a New Strip-Based Blood Glucose System for Point-of-Care Testing in Critical and Non-Critical Care Settings
Journal of Diabetes Science and Technology · 2025
DOI: 10.1177/19322968251325873 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Journal of Diabetes Science and Technology, 2025. Verfasst von Michael Goodman und 25 weiteren Autor:innen. Beteiligte Einrichtung: Department of Surgery, University of Cincinnati, Cincinnati, OH, USA. Hauptthemen laut MEDLINE: Blood Glucose, Point-of-Care Testing, Point-of-Care Systems, Critical Care. Die Arbeit nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Roche Diagnostics.
Abstract (Originalsprache)
Background
Evaluation of the performance of Cobas ® Pulse (Roche Diagnostics GmbH, Mannheim, Germany), a new blood glucose (BG) monitoring system (BGMS; referred to here as BGMSA) intended for point-of-care testing using samples from patients in diverse clinical settings by intended point-of-care test operators.
Methods
Arterial, capillary, venous, or heel stick whole blood (WB) samples from patients in non-critical and critical care settings were collected and analyzed using BGMSA and the Nova StatStrip ® BGMS (Nova Biomedical, Waltham, MA; referred to here as BGMSB), and a hexokinase comparator (Cobas 6000 Analyzer Series; Roche Diagnostics GmbH). The blood glucose measurement accuracy was assessed by the Food and Drug Administration guidance criteria.
Results
Two studies are presented. In the first, 2678 samples (622 arterial, 706 capillary, 1203 venous, 147 heel stick) were collected from 1577 patients in 14 US and three European sites. All accuracy criteria were met for arterial and venous samples considering all data combined. BGMSA showed better accuracy than BGMSB for arterial, venous, and heel stick, and similar results to BGMSB for capillary WB vs venous comparator. No endogenous interference from pO 2 , hematocrit, and sodium was identified. BGMSA was also accurate when analyzing contrived samples used to show accuracy over a wide range of glucose concentrations, alone and when combined with extreme hematocrit. In the second study, 117 capillary fingerstick samples collected at one US site were measured on both BGMSA and the comparator, and all accuracy criteria were met. No clinically significant medical risks were observed via Diabetes Technology Society Error Grids.
Conclusions
BGMSA was effective for determining blood glucose in venous, arterial, neonatal arterial, neonatal heel stick, and capillary WB samples.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Surgery, University of Cincinnati, Cincinnati, OH, USA.
- Department of Anesthesiology and Critical Care Medicine, University Hospital, Leipzig, Germany.
- NB Research, Inc., Indianapolis, IN, USA.
- Roche Molecular Systems, Inc., Indianapolis, IN, USA.
- Roche Diagnostics International AG, Rotkreuz, Switzerland.
- Department of Pathology, Johns Hopkins University School of Medicine, Baltimore, MD, USA.
ORCID-Profile: Michael Goodman, James H Nichols, Nam K Tran, Alan H B Wu
Förderung
- Roche Diagnostics
Zitieren
APA
Goodman, M., Bercker, S., Breitenbeck, N., Canada-Vilalta, C., Canepa, D. D., Clarke, W. A., Farnsworth, C. W., Ganser, E. O., Geerts, N., Goldsmith, B., Headden, G., Hoppensteadt, D., Klapperich, B., Matika, R., Miles, G., Mukhopadhyay, S., Nichols, J. H., O'Dell, R. M., Rafique, Z., …, & Karon, B. S. (2025). Multicenter Evaluation of a New Strip-Based Blood Glucose System for Point-of-Care Testing in Critical and Non-Critical Care Settings. Journal of Diabetes Science and Technology. https://doi.org/10.1177/19322968251325873
Vancouver
Goodman M, Bercker S, Breitenbeck N, Canada-Vilalta C, Canepa DD, Clarke WA, et al. Multicenter Evaluation of a New Strip-Based Blood Glucose System for Point-of-Care Testing in Critical and Non-Critical Care Settings. Journal of Diabetes Science and Technology. 2025. doi:10.1177/19322968251325873
BibTeX
@article{Goodman2025,
author = {Goodman, Michael and Bercker, Sven and Breitenbeck, Nancy and Canada-Vilalta, Cristina and Canepa, Daisy D and Clarke, William A and Farnsworth, Christopher W and Ganser, Edward O and Geerts, Nienke and Goldsmith, Barbara and Headden, Gary and Hoppensteadt, Debra and Klapperich, Birgit and Matika, Ryan and Miles, Gabrielle and Mukhopadhyay, Sagori and Nichols, James H and O'Dell, Rebecca M and Rafique, Zubaid and Schuetzenmeister, Andre and Slingerland, Robbert and Stankiewicz, Ann F and Swanson, Jonathan R and Tran, Nam K and Wu, Alan H B and Karon, Brad S},
title = {Multicenter Evaluation of a New Strip-Based Blood Glucose System for Point-of-Care Testing in Critical and Non-Critical Care Settings},
journal = {Journal of Diabetes Science and Technology},
year = {2025},
doi = {10.1177/19322968251325873}
}
EUVIMED