Point-of-Care Testing Glucometer in Rapid Assessment Settings: Comparison of 2 Network-Capable Devices in a Referral Hospital

Giacomo Moretti iD, Giorgia Degni, Camilla Lamiano, Eugenia Tulli, Andrea Urbani

The Journal of Applied Laboratory Medicine · 2025

DOI: 10.1093/jalm/jfaf048 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift The Journal of Applied Laboratory Medicine, 2025. Verfasst von Giacomo Moretti und 4 weiteren Autor:innen. Beteiligte Einrichtung: Clinical Chemistry, Biochemistry and Molecular Biology Operations (UOC), Fondazione Policlinico Universitario A. Gemelli IRCCS, Rome, Italy. Hauptthemen laut MEDLINE: Blood Glucose, Point-of-Care Testing, Blood Glucose Self-Monitoring, Point-of-Care Systems. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 11 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Point-of-care testing (POCT) glucose meters are essential for rapid glucose monitoring. This study aimed to evaluate the analytical performance of 2 novel network-capable POCT glucometers, CobasPulse (Roche Diagnostic) and StatStrip (Nova Biomedical), in comparison to the Atellica CH 930 Analyzer (Siemens Healthcare), a central laboratory clinical chemistry analyzer, as a reference method.

Methods

A cohort of 150 patients’ venous whole blood samples were analyzed. Method comparison was performed using Passing–Bablok regression and Bland–Altman plots. Precision studies were conducted using commercial controls, with assessment of within-run and between-run imprecision by experienced laboratory technicians. Performance was assessed against FDA 2020 benchmarks.

Results

Passing–Bablok regression showed accurate overlap for both POCT devices. CobasPulse exhibited a bias of −0.17 mmol/L and Pearson r of 0.982, while StatStrip showed a bias of −0.35 mmol/L and Pearson r of 0.959. Within-run CV for Cobas Pulse (3.3 mmol/L control) was 2.4%, and for StatStrip (3.44 mmol/L control) was 5.2%. Between-run CVs were 2.1% for Cobas Pulse and 3.2% for StatStrip at comparable glucose concentrations.

Conclusions

Both CobasPulse and StatStrip demonstrated acceptable concordance with the Atellica CH 930 Analyzer. While Cobas Pulse showed slightly better agreement and precision, both devices are suitable for glucose measurements in healthcare settings, providing reliable results and reduced turnaround time, supporting timely clinical decisions for rapid glycemic assessment.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Blood GlucosePoint-of-Care TestingBlood Glucose Self-MonitoringPoint-of-Care SystemsHumansReproducibility of ResultsMaleFemaleMiddle AgedReferral and Consultation

Substanzen

Blood Glucose

Autor:innen und Institutionen

  • Clinical Chemistry, Biochemistry and Molecular Biology Operations (UOC), Fondazione Policlinico Universitario A. Gemelli IRCCS, Rome, Italy.
  • Department of Basic Biotechnological Sciences, Intensive Care and Perioperative Clinics Research, Catholic University of the Sacred Heart, Rome, Italy.

ORCID-Profile: Giacomo Moretti

Zitieren

APA

Moretti, G., Degni, G., Lamiano, C., Tulli, E., & Urbani, A. (2025). Point-of-Care Testing Glucometer in Rapid Assessment Settings: Comparison of 2 Network-Capable Devices in a Referral Hospital. The Journal of Applied Laboratory Medicine. https://doi.org/10.1093/jalm/jfaf048

Vancouver

Moretti G, Degni G, Lamiano C, Tulli E, Urbani A. Point-of-Care Testing Glucometer in Rapid Assessment Settings: Comparison of 2 Network-Capable Devices in a Referral Hospital. The Journal of Applied Laboratory Medicine. 2025. doi:10.1093/jalm/jfaf048

BibTeX
@article{Moretti2025,
  author = {Moretti, Giacomo and Degni, Giorgia and Lamiano, Camilla and Tulli, Eugenia and Urbani, Andrea},
  title = {Point-of-Care Testing Glucometer in Rapid Assessment Settings: Comparison of 2 Network-Capable Devices in a Referral Hospital},
  journal = {The Journal of Applied Laboratory Medicine},
  year = {2025},
  doi = {10.1093/jalm/jfaf048}
}
Kostenlos starten →