Performance of Three Continuous Glucose Monitoring Systems in Adults With Type 1 Diabetes

Leistung von drei kontinuierlichen Glukosemesssystemen bei Erwachsenen mit Typ-1-Diabetes

Manuel Eichenlaub iD, Delia Waldenmaier iD, Stephanie Wehrstedt iD, Stefan Pleus iD, Manuela Link, Nina Jendrike, Sükrü Öter iD, Cornelia Haug, Maren Schinz, Vincent Braunack-Mayer, Regula Schneider, Derek Brandt und 1 weitere

Journal of Diabetes Science and Technology · 2026-07 · 20 (4) · 1131-1140

DOI: 10.1177/19322968251315459 ↗

Auf einen Blick

Bei 24 Erwachsenen mit Typ-1-Diabetes wurden FreeStyle Libre 3, Dexcom G7 und Medtronic Simplera parallel geprüft. Die Genauigkeit variierte je nach Vergleichsmethode; FL3 und DG7 waren insgesamt tendenziell genauer als MSP.

Fragestellung

Wie unterscheiden sich die Leistungen von FreeStyle Libre 3, Dexcom G7 und Medtronic Simplera gegenüber verschiedenen Vergleichsmethoden und in relevanten glykämischen Situationen?

Methode

24 Erwachsene mit Typ-1-Diabetes trugen bis zu 15 Tage je einen Sensor der 3 CGM-Systeme parallel. Verglichen wurde mit YSI 2300, Cobas Integra und Contour Next; zusätzlich wurden Hyper- und Hypoglykämie-Szenarien induziert.

Ergebnisse

Gegenüber YSI lagen die MARDs bei 11,6 % (FL3), 12,0 % (DG7) und 11,6 % (MSP). Gegenüber INT: 9,5 %, 9,9 % und 13,9 %. Gegenüber CNX: 9,7 %, 10,1 % und 16,6 %. FL3 und DG7 waren in normo- und hyperglykämischen Bereichen besser, MSP im hypoglykämischen Bereich.

Einschränkungen

Die Genauigkeit hing von der Vergleichsmethode ab; alle Systeme zeigten eine niedrigere Genauigkeit als in früheren Studien.

Schlussfolgerung der Autor:innen

Die Leistungen aller CGM-Systeme variierten je nach Vergleichsmethode. FL3 und DG7 waren über die Vergleiche hinweg tendenziell genauer als MSP.

Kernaussagen

  • 24 Erwachsene mit Typ-1-Diabetes
  • MARD gegenüber YSI: 11,6 % / 12,0 % / 11,6 %
  • FL3 und DG7 insgesamt tendenziell genauer als MSP

Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Diabetes Science and Technology, online erschienen am 04.02.2025. Verfasst von Manuel Eichenlaub und 12 weiteren Autor:innen. Beteiligte Einrichtung: Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany. Hauptthemen laut MEDLINE: Diabetes Mellitus, Type 1, Continuous Glucose Monitoring, Blood Glucose, Blood Glucose Self-Monitoring. Die Arbeit wurde bisher 50-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: BIONIME Corporation, Roche Diabetes Care GmbH, iSENS, Inc..

Auf einen BlickAbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The performance of continuous glucose monitoring (CGM) systems is difficult to compare due to different study designs and a lack of head-to-head studies. This study evaluated the performance of FreeStyle Libre 3 (FL3), Dexcom G7 (DG7), and Medtronic Simplera (MSP) against different comparator methods and during clinically relevant glycemic scenarios.

Method

Twenty-four adult participants with type 1 diabetes mellitus wore one sensor of each CGM system in parallel for up to 15 days. Sensors of DG7 and MSP were exchanged on days 5 and 8, respectively. Three 7-hour sessions with 15-minute comparator blood glucose–level measurements using YSI 2300 (YSI, venous), Cobas Integra (INT, venous), and Contour Next (CNX, capillary) were conducted on days 2, 5, and 15. Simultaneously, glucose-level excursions with transient hyperglycemia and hypoglycemia were induced according to a recently published testing procedure. The accuracy was evaluated using various metrics, including mean absolute relative differences (MARDs).

Results

Compared with YSI data, the MARDs of FL3, DG7, and MSP were 11.6%, 12.0%, and 11.6%, respectively. Relative to the INT data, the corresponding MARDs were 9.5%, 9.9%, and 13.9%, respectively, and compared with CNX data, MARDs were 9.7%, 10.1%, and 16.6%, respectively. Both FL3 and DG7 showed better accuracy in the normoglycemic and hyperglycemic range, while MSP performed better in the hypoglycemic range.

Conclusions

Performance results of all CGM systems varied depending on the comparator method. However, across comparators FL3 and DG7 tended to be more accurate compared with MSP. All CGM systems showed a lower accuracy compared with previous studies, emphasizing the need for comprehensive study design guidelines.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Diabetes Mellitus, Type 1Continuous Glucose MonitoringBlood GlucoseBlood Glucose Self-MonitoringHumansAdultFemaleMaleMiddle AgedReproducibility of ResultsYoung Adult

Stichwörter der Autor:innen

CG-DIVACGM performanceaccuracycomparator methoddynamic glucose regions

Substanzen

Blood Glucose

Autor:innen und Institutionen

  • Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Ulm, Germany.
  • Diabetes Center Berne, Bern, Switzerland.

ORCID-Profile: Manuel Eichenlaub, Delia Waldenmaier, Stephanie Wehrstedt, Stefan Pleus, Sükrü Öter, Guido Freckmann

Förderung

  • BIONIME Corporation
  • Roche Diabetes Care GmbH
  • iSENS, Inc.
  • Ascensia Holdings AG
  • Diabetes Center Berne

Zitieren

APA

Eichenlaub, M., Waldenmaier, D., Wehrstedt, S., Pleus, S., Link, M., Jendrike, N., Öter, S., Haug, C., Schinz, M., Braunack-Mayer, V., Schneider, R., Brandt, D., & Freckmann, G. (2026). Performance of Three Continuous Glucose Monitoring Systems in Adults With Type 1 Diabetes. Journal of Diabetes Science and Technology, 20(4), 1131-1140. https://doi.org/10.1177/19322968251315459

Vancouver

Eichenlaub M, Waldenmaier D, Wehrstedt S, Pleus S, Link M, Jendrike N, et al. Performance of Three Continuous Glucose Monitoring Systems in Adults With Type 1 Diabetes. Journal of Diabetes Science and Technology. 2026;20(4):1131-1140. doi:10.1177/19322968251315459

BibTeX
@article{Eichenlaub2026,
  author = {Eichenlaub, Manuel and Waldenmaier, Delia and Wehrstedt, Stephanie and Pleus, Stefan and Link, Manuela and Jendrike, Nina and Öter, Sükrü and Haug, Cornelia and Schinz, Maren and Braunack-Mayer, Vincent and Schneider, Regula and Brandt, Derek and Freckmann, Guido},
  title = {Performance of Three Continuous Glucose Monitoring Systems in Adults With Type 1 Diabetes},
  journal = {Journal of Diabetes Science and Technology},
  year = {2026},
  volume = {20},
  number = {4},
  pages = {1131--1140},
  doi = {10.1177/19322968251315459}
}
Kostenlos starten →