Highly Miniaturized, Low-Power CMOS ASIC Chip for Long-Term Continuous Glucose Monitoring
Journal of Diabetes Science and Technology · 2024
DOI: 10.1177/19322968231153419 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Diabetes Science and Technology, 2024. Verfasst von Raja Hari Gudlavalleti und 8 weiteren Autor:innen. Beteiligte Einrichtung: University of Connecticut, Storrs, CT, USA. Hauptthemen laut MEDLINE: Blood Glucose, Blood Glucose Self-Monitoring, Semiconductors, Biosensing Techniques. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: JDRF, US Army CDMRP, Leona M. and Harry B. Helmsley Charitable Trust.
Abstract (Originalsprache)
Background
The objective of this work is to develop a highly miniaturized, low-power, biosensing platform for continuous glucose monitoring (CGM). This platform is based on an application-specific integrated circuit (ASIC) chip that interfaces with an amperometric glucose-sensing element. To reduce both size and power requirements, this custom ASIC chip was implemented using 65-nm complementary metal oxide semiconductor (CMOS) technology node. Interfacing this chip to a frequency-counting microprocessor with storage capabilities, a miniaturized transcutaneous CGM system can be constructed for small laboratory animals, with long battery life.
Method
A 0.45 mm × 1.12 mm custom ASIC chip was first designed and implemented using the Taiwan Semiconductor Manufacturing Company (TSMC) 65-nm CMOS technology node. This ASIC chip was then interfaced with a multi-layer amperometric glucose-sensing element and a frequency-counting microprocessor with storage capabilities. Variation in glucose levels generates a linear increase in frequency response of this ASIC chip. In vivo experiments were conducted in healthy Sprague Dawley rats.
Results
This highly miniaturized, 65-nm custom ASIC chip has an overall power consumption of circa 36 µW. In vitro testing shows that this ASIC chip produces a linear ( R 2 = 99.5) frequency response to varying glucose levels (from 2 to 25 mM), with a sensitivity of 1278 Hz/mM. In vivo testing in unrestrained healthy rats demonstrated long-term CGM (six days/per charge) with rapid glucose response to glycemic variations induced by isoflurane anesthesia and tail vein injection.
Conclusions
The miniature footprint of the biosensor platform, together with its low-power consumption, renders this CMOS ASIC chip a versatile platform for a variety of highly miniaturized devices, intended to improve the quality of life of patients with type 1 and type 2 diabetes.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- University of Connecticut, Storrs, CT, USA.
- Biorasis Inc., Storrs, CT, USA.
ORCID-Profile: Raja Hari Gudlavalleti, Diane Burgess
Förderung
- JDRF
- US Army CDMRP
- Leona M. and Harry B. Helmsley Charitable Trust
Zitieren
APA
Gudlavalleti, R. H., Xi, X., Legassey, A., Chan, P. Y., Li, J., Burgess, D., Giardina, C., Papadimitrakopoulos, F., & Jain, F. (2024). Highly Miniaturized, Low-Power CMOS ASIC Chip for Long-Term Continuous Glucose Monitoring. Journal of Diabetes Science and Technology. https://doi.org/10.1177/19322968231153419
Vancouver
Gudlavalleti RH, Xi X, Legassey A, Chan PY, Li J, Burgess D, et al. Highly Miniaturized, Low-Power CMOS ASIC Chip for Long-Term Continuous Glucose Monitoring. Journal of Diabetes Science and Technology. 2024. doi:10.1177/19322968231153419
BibTeX
@article{Gudlavalleti2024,
author = {Gudlavalleti, Raja Hari and Xi, Xiangyi and Legassey, Allen and Chan, Pik-Yiu and Li, Jin and Burgess, Diane and Giardina, Charles and Papadimitrakopoulos, Fotios and Jain, Faquir},
title = {Highly Miniaturized, Low-Power CMOS ASIC Chip for Long-Term Continuous Glucose Monitoring},
journal = {Journal of Diabetes Science and Technology},
year = {2024},
doi = {10.1177/19322968231153419}
}
EUVIMED