EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Angiographic Safety and Efficacy of the ReSolv Flow-Diverting Stent in a Rabbit Model

Rosalie Ea Morrish iD, Alec T Chunta iD, Brooke L Belanger iD, Paige M Croney, M Suheel Abdul Salam, Crista Thompson, Muneer Eesa, John H Wong, Alim P Mitha

Interventional Neuroradiology · 2026

DOI: 10.1177/15910199241260896 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Interventional Neuroradiology, 2026. Verfasst von Rosalie Ea Morrish und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Neurosciences, University of Calgary, Calgary, Alberta, Canada. Hauptthemen laut MEDLINE: Stents, Intracranial Aneurysm, Cerebral Angiography, Endovascular Procedures. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Research Council of Canada Industrial Research Assistance Program, Evolve2Innovate, NSERC Brain CREATE.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Bioresorbable polymer-based flow-diverting stents have potential benefits over existing metal devices. This study aimed to evaluate the safety and efficacy of the novel ReSolv device, which is a primarily polymer-based flow-diverting stent, using the in vivo rabbit sidewall saccular aneurysm model.

Methods

ReSolv stents were deployed in 14 New Zealand White rabbits that had undergone aneurysm creation procedures. Animals were allocated to follow-up time points of 1, 3, 6, 9, 12, 16, or 18 months. Angiographic images were evaluated by an independent neurointerventionalist blinded to follow-up time points for (1) in-stent stenosis, (2) parent vessel and jailed side branch patency, (3) wall apposition, and (4) aneurysm occlusion using the Raymond-Roy Occlusion Classification (RROC), O’Kelly Marotta grading scale, and the 4F flow diversion predictive score. Primary efficacy outcome was defined as RROC Class I or II.

Results

At a median follow-up time of 7.5 months, parent vessel (14/14) and jailed side (33/33) branches were patent in all cases. There was no development of thrombus on the stent or cases of significant in-stent stenosis, and all stents had good wall apposition. Adequate occlusion was found in 85.7% ( n = 12) of animals, including an RROC Class I in 64.3% ( n = 9) and RROC Class II in 21.4% ( n = 3).

Conclusions

The ReSolv stent shows encouraging angiographic safety and efficacy outcomes after placement in a rabbit sidewall saccular aneurysm model. Longer term studies are ongoing to determine eventual fate of the aneurysm, parent vessel, and jailed side branches after absorption of the polymer component of the stent.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

StentsIntracranial AneurysmCerebral AngiographyEndovascular ProceduresAnimalsRabbitsDisease Models, AnimalProsthesis DesignTreatment Outcome

Stichwörter der Autor:innen

aneurysmefficacyendovascularflow-diverting stentocclusionsafety

Autor:innen und Institutionen

  • Department of Clinical Neurosciences, University of Calgary, Calgary, Alberta, Canada.
  • Hotchkiss Brain Institute, University of Calgary, Calgary, Alberta, Canada.
  • Department of Biomedical Engineering, University of Calgary, Calgary, Alberta, Canada.
  • Section of Neurosurgery, University of Manitoba, Winnipeg, Manitoba, Canada.
  • Fluid Biomed, Calgary, Alberta, Canada.

ORCID-Profile: Rosalie Ea Morrish, Alec T Chunta, Brooke L Belanger

Förderung

  • National Research Council of Canada Industrial Research Assistance Program
  • Evolve2Innovate
  • NSERC Brain CREATE

Zitieren

APA

Morrish, R. E., Chunta, A. T., Belanger, B. L., Croney, P. M., Salam, M. S. A., Thompson, C., Eesa, M., Wong, J. H., & Mitha, A. P. (2026). Angiographic Safety and Efficacy of the ReSolv Flow-Diverting Stent in a Rabbit Model. Interventional Neuroradiology. https://doi.org/10.1177/15910199241260896

Vancouver

Morrish RE, Chunta AT, Belanger BL, Croney PM, Salam MSA, Thompson C, et al. Angiographic Safety and Efficacy of the ReSolv Flow-Diverting Stent in a Rabbit Model. Interventional Neuroradiology. 2026. doi:10.1177/15910199241260896

BibTeX
@article{Morrish2026,
  author = {Morrish, Rosalie Ea and Chunta, Alec T and Belanger, Brooke L and Croney, Paige M and Salam, M Suheel Abdul and Thompson, Crista and Eesa, Muneer and Wong, John H and Mitha, Alim P},
  title = {Angiographic Safety and Efficacy of the ReSolv Flow-Diverting Stent in a Rabbit Model},
  journal = {Interventional Neuroradiology},
  year = {2026},
  doi = {10.1177/15910199241260896}
}
Kostenlos starten →