Ikelos-RWA. Validation of an Automatic Tool to Quantify REM Sleep Without Atonia
Clinical EEG and Neuroscience · 2024
DOI: 10.1177/15500594231175320 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Clinical EEG and Neuroscience, 2024. Verfasst von Alexandra Papakonstantinou und 4 weiteren Autor:innen. Beteiligte Einrichtung: Sleep Research & Clinical Chronobiology, Institute of Physiology, Charité - Universitätsmedizin Berlin, Corporate Member of Freien Universität Berlin…. Hauptthemen laut MEDLINE: Polysomnography, REM Sleep Behavior Disorder, Algorithms, Sleep, REM. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Study Objectives. To present and evaluate an automatic scoring algorithm for quantification of REM-sleep without atonia (RWA) in patients with REM-sleep behaviour disorder (RBD) based on a generally accepted, well-validated visual scoring method, (“Montreal” phasic and tonic) and a recently developed, concise scoring method (Ikelos-RWA).
Methods
Video-polysomnographies of 20 RBD patients (68.2 ± 7.2 years) and 20 control patients with periodic limb movement disorder (65.9 ± 6.7 years) were retrospectively analysed. RWA was estimated from chin electromyogram during REM-sleep. Visual and automated RWA scorings were correlated, and agreement ( a ) and Cohen's Kappa ( k ) calculated for 1735 minutes of REM-sleep of the RBD patients. Discrimination performance was evaluated with receiver operating characteristic (ROC) analysis. The algorithm was then applied on the polysomnographies of a cohort of 232 RBD patients (total analysed REM-sleep: 17,219 minutes) and evaluated, while correlating the different output parameters.
Results
Visual and computer-derived RWA scorings correlated significantly (tonic Montreal: r TM = 0.77; phasic Montreal: r PM = 0.78; Ikelos-RWA: r I = 0.97; all p < 0.001) and showed good to excellent Kappa coefficients ( k TM = 0.71; k PM = 0.79; k I = 0.77). The ROC analysis showed high sensitivities (95%-100%) and specificities (84%-95%) at the optimal operation points, with area under the curve (AUC) of 0.98, indicating high discriminating capacity. The automatic RWA scorings of 232 patients correlated significantly ( r TM{I} = 0.95; r PM{I} = 0.91, p < 0.0001).
Conclusions
The presented algorithm is an easy-to-use and valid tool for automatic RWA scoring in patients with RBD and may prove effective for general use being publicly available.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Sleep Research & Clinical Chronobiology, Institute of Physiology, Charité - Universitätsmedizin Berlin, Corporate Member of Freien Universität Berlin, Humboldt-Universität zu Berlin, and Berlin Institute of Health, Berl…
- Clinic for Sleep- and Chronomedicine, St. Hedwig-Krankenhaus, Berlin, Germany.
- Department of Ophthalmology, University Medical Center Goettingen, Göttingen, Germany.
ORCID-Profile: Frederik Bes
Zitieren
APA
Papakonstantinou, A., Klemming, J., Haberecht, M., Kunz, D., & Bes, F. (2024). Ikelos-RWA. Validation of an Automatic Tool to Quantify REM Sleep Without Atonia. Clinical EEG and Neuroscience. https://doi.org/10.1177/15500594231175320
Vancouver
Papakonstantinou A, Klemming J, Haberecht M, Kunz D, Bes F. Ikelos-RWA. Validation of an Automatic Tool to Quantify REM Sleep Without Atonia. Clinical EEG and Neuroscience. 2024. doi:10.1177/15500594231175320
BibTeX
@article{Papakonstantinou2024,
author = {Papakonstantinou, Alexandra and Klemming, Jannis and Haberecht, Martin and Kunz, Dieter and Bes, Frederik},
title = {Ikelos-RWA. Validation of an Automatic Tool to Quantify REM Sleep Without Atonia},
journal = {Clinical EEG and Neuroscience},
year = {2024},
doi = {10.1177/15500594231175320}
}
EUVIMED