Accuracy of surface-guided radiotherapy-based positioning and surface tracking in breast radiotherapy using a thermo-optical camera system

Hiroki Katayama, Yosuke Takahashi, Motonori Kitaoka, Hiroki Kawasaki, Takashi Tanii, Yayoi Taniguchi, Masato Tsuzuki

Physical and Engineering Sciences in Medicine · 2026

DOI: 10.1007/s13246-025-01649-4 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physical and Engineering Sciences in Medicine, 2026. Verfasst von Hiroki Katayama und 6 weiteren Autor:innen. Beteiligte Einrichtung: Division of Clinical Radiology, Department of Medical Technology, Kagawa University Hospital, 1750-1 Ikenobe, Miki-cho, Kita-gun, Kagawa, 761-0793, J…. Hauptthemen laut MEDLINE: Breast Neoplasms, Radiotherapy, Image-Guided, Patient Positioning. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 25 Literaturangaben. Gefördert von: Japan Society for the Promotion of Science.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study aimed to evaluate the accuracy of surface-guided radiotherapy (SGRT)-based positioning and surface tracking in breast radiotherapy using the ExacTrac Dynamic (ETD). We retrospectively evaluated 16 patients who underwent breast cancer treatment. All patients were positioned using SGRT with ETD under free-breathing conditions. Orthogonal kilovoltage (kV) image acquisition was then acquired using the linear accelerators' kV imagers and registered to digitally reconstructed radiographs. For surface tracking, vertical breast surface motion waveforms obtained with ETD were compared with those captured using an electronic portal imaging device (EPID) during beam delivery. The agreement between ETD and EPID in terms of vertical breast surface displacement was evaluated using the cross-correlation coefficient. For patient positioning, the mean ± standard deviation (SD) of X-ray shift values (mm) were - 2.8 ± 2.6 (vertical), - 2.2 ± 4.5 (longitudinal), and 0.3 ± 2.1(lateral). The percentage of X-ray shift < 5 mm was 82% (vertical), 66% (longitudinal), and 92% (lateral). The setup margins (mm) were 5.4 (vertical), 9.0 (longitudinal), and 3.8 (lateral). For surface tracking, the mean ± SD of the cross-correlation coefficient was 0.93 ± 0.02, indicating a high correlation between the ETD and EPID waveforms. The mean amplitude difference between waveforms obtained by ETD and EPID was 0.46 mm. SGRT-based positioning with ETD may result in errors exceeding 5 mm relative to the treatment planning position in the longitudinal direction; combining it with image-guided radiotherapy is therefore recommended. Surface tracking with ETD demonstrated high tracking accuracy in tracking human body surface temperature distribution.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Breast NeoplasmsRadiotherapy, Image-GuidedPatient PositioningHumansFemaleMiddle AgedRetrospective StudiesAgedSurface PropertiesRadiotherapy Planning, Computer-Assisted

Stichwörter der Autor:innen

Breast cancerExacTrac dynamicSGRTSGRT positioningSurface tracking

Autor:innen und Institutionen

  • Division of Clinical Radiology, Department of Medical Technology, Kagawa University Hospital, 1750-1 Ikenobe, Miki-cho, Kita-gun, Kagawa, 761-0793, Japan. katayama.hiroki@kagawa-u.ac.jp.
  • Division of Clinical Radiology, Department of Medical Technology, Kagawa University Hospital, 1750-1 Ikenobe, Miki-cho, Kita-gun, Kagawa, 761-0793, Japan.

Förderung

  • Japan Society for the Promotion of Science

Zitieren

APA

Katayama, H., Takahashi, Y., Kitaoka, M., Kawasaki, H., Tanii, T., Taniguchi, Y., & Tsuzuki, M. (2026). Accuracy of surface-guided radiotherapy-based positioning and surface tracking in breast radiotherapy using a thermo-optical camera system. Physical and Engineering Sciences in Medicine. https://doi.org/10.1007/s13246-025-01649-4

Vancouver

Katayama H, Takahashi Y, Kitaoka M, Kawasaki H, Tanii T, Taniguchi Y, et al. Accuracy of surface-guided radiotherapy-based positioning and surface tracking in breast radiotherapy using a thermo-optical camera system. Physical and Engineering Sciences in Medicine. 2026. doi:10.1007/s13246-025-01649-4

BibTeX
@article{Katayama2026,
  author = {Katayama, Hiroki and Takahashi, Yosuke and Kitaoka, Motonori and Kawasaki, Hiroki and Tanii, Takashi and Taniguchi, Yayoi and Tsuzuki, Masato},
  title = {Accuracy of surface-guided radiotherapy-based positioning and surface tracking in breast radiotherapy using a thermo-optical camera system},
  journal = {Physical and Engineering Sciences in Medicine},
  year = {2026},
  doi = {10.1007/s13246-025-01649-4}
}
Kostenlos starten →