Impact of 1.5 T Magnetic Field on Treatment Plan Quality in MR-Guided Radiotherapy: Typical Phantom Test Cases

Lingling Yan, Yingjie Xu, Jianrong Dai iD

Technology in Cancer Research & Treatment · 2024

DOI: 10.1177/15330338241272038 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Technology in Cancer Research & Treatment, 2024. Verfasst von Lingling Yan und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sci…. Hauptthemen laut MEDLINE: Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted, Radiotherapy Dosage, Radiotherapy, Image-Guided. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Guangzhou Shared Future Charitable Foundation "Dengfeng" research project of precision radiotherapy on cancer, National High Level Hospital Clinical Research Funding.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This study aims to investigate the influence of the magnetic field on treatment plan quality using typical phantom test cases, which encompass a circle target test case, AAPM TG119 test cases (prostate, head-and-neck, C-shape, multi-target test cases), and a lung test case. Materials and Methods For the typical phantom test cases, two plans were formulated. The first plan underwent optimization in the presence of a 1.5 Tesla magnetic field (1.5 T plan). The second plan was re-optimized without a magnetic field (0 T plan), utilizing the same optimization conditions as the first plan. The two plans were compared based on various parameters, including con­formity index (CI), homogeneity index (HI), fit index (FI) and dose coverage of the planning target volume (PTV), dose delivered to organs at risk (OARs) and normal tissue (NT), monitor unit (MU). A plan-quality metric (PQM) scoring procedure was employed. For the 1.5 T plans, dose verifications were performed using an MR-compatible ArcCHECK phantom.

Results

A smaller dose influence of the magnetic field was found for the circle target, prostate, head-and-neck, and C-shape test cases, compared with the multi-target and lung test cases. In the multi-target test case, the significant dose influence was on the inferior PTV, followed by the superior PTV. There was a relatively large dose influence on the PTV and OARs for lung test case. No statistically significant differences in PQM and MUs were observed. For the 1.5 T plans, gamma passing rates were all higher than 95% with criteria of 2 mm/3% and 2 mm/2%.

Conclusion

The presence of a 1.5 T magnetic field had a relatively large impact on dose parameters in the multi-target and lung test cases compared with other test cases. However, there were no significant influences on the plan-quality metric, MU and dose accuracy for all test cases.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingRadiotherapy Planning, Computer-AssistedRadiotherapy DosageRadiotherapy, Image-GuidedMagnetic FieldsMagnetic Resonance ImagingHumansOrgans at RiskNeoplasmsMaleRadiotherapy, Intensity-ModulatedProstatic Neoplasms

Stichwörter der Autor:innen

1.5 T magnetic fieldMR-guided radiotherapygamma passing ratemonitor unitplan-quality metrictreatment plan qualitytypical phantom test case

Autor:innen und Institutionen

  • Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.

ORCID-Profile: Jianrong Dai

Förderung

  • Guangzhou Shared Future Charitable Foundation "Dengfeng" research project of precision radiotherapy on cancer
  • National High Level Hospital Clinical Research Funding

Zitieren

APA

Yan, L., Xu, Y., & Dai, J. (2024). Impact of 1.5 T Magnetic Field on Treatment Plan Quality in MR-Guided Radiotherapy: Typical Phantom Test Cases. Technology in Cancer Research & Treatment. https://doi.org/10.1177/15330338241272038

Vancouver

Yan L, Xu Y, Dai J. Impact of 1.5 T Magnetic Field on Treatment Plan Quality in MR-Guided Radiotherapy: Typical Phantom Test Cases. Technology in Cancer Research & Treatment. 2024. doi:10.1177/15330338241272038

BibTeX
@article{Yan2024,
  author = {Yan, Lingling and Xu, Yingjie and Dai, Jianrong},
  title = {Impact of 1.5 T Magnetic Field on Treatment Plan Quality in MR-Guided Radiotherapy: Typical Phantom Test Cases},
  journal = {Technology in Cancer Research & Treatment},
  year = {2024},
  doi = {10.1177/15330338241272038}
}
Kostenlos starten →