The Varying Benefits of Two Years Automated Insulin Delivery System Use on Continuous Glucose Monitoring Profiles in Children and Adolescents with Type 1 Diabetes
Diabetes Technology & Therapeutics · 21.11.2026
DOI: 10.1177/15209156261507763 ↗Kurz gefasst
Fachartikel in der Zeitschrift Diabetes Technology & Therapeutics, online erschienen am 21.11.2026. Verfasst von Jantje Weiskorn und 7 weiteren Autor:innen. Die Arbeit nennt 31 Literaturangaben.
Themen: Diabetes mellitus, Kinderkrankenpflege
Abstract (Originalsprache)
Background
Automated insulin delivery (AID) systems are recommended for youths with type 1 diabetes, but real-world longitudinal data remains limited despite increasing use. This study evaluates the effect of 2 years of AID use on continuous glucose monitoring (CGM) profiles and body mass index (BMI) development.
Methods
Data from individuals with type 1 diabetes aged 3 months, and AID use of >3 months from the Diabetes Prospective Follow-up (DPV) Registry between 2019 and 2025 were analyzed. CGM and HbA1c were required at the last documented time in range (TIR) (±90 days; follow-up) and 2 years earlier (±90 days; baseline). Changes were investigated using adjusted regression models with repeated measurements.
Findings
Data from 2486 participants (baseline: median age: 11.9 years, diabetes duration: 5.0 years, 53% male) were analyzed over ∼23 months. The largest gains occurred in time in tight range (TiTR) (42.4% [95% confidence interval: 41.0–43.7] to 50.1% [48.8–51.4]; P < 0.001) and coefficient of variation (CV) (32.9 [32.0–33.7] to 29.5 [28.7–30.2]; P < 0.001). HbA1c (7.3% [7.2–7.3] to 7.2 [7.1–7.2]; P < 0.001), TIR (64.5% [63.9–65.1] to 66.7% [66.1–67.3]; P < 0.001), and mean sensor glucose (SG) (161.0 mg/dL [158.4–163.5] to 155.5 mg/dL [153.1–157.9]; P = 0.001) improved statistically, but below clinical relevance thresholds. Time below range (TBR) < 70 mg/dL increased (2.7% [2.6–2.9] to 3.2% [3.1–3.4]; P < 0.001), TBR < 54 mg/dL remained stable (1.7%–1.8%; P = 0.516), and severe hypoglycemia decreased (6.4–4.5/100 patient-years; P < 0.001). No changes were seen in children
Zitieren
APA
Weiskorn, J., Auzanneau, M., Burckhardt, M., Blauensteiner, N., Schierloh, U., Mayer, I., Schneidewind, J., & Lanzinger, S. (2026). The Varying Benefits of Two Years Automated Insulin Delivery System Use on Continuous Glucose Monitoring Profiles in Children and Adolescents with Type 1 Diabetes. Diabetes Technology & Therapeutics. https://doi.org/10.1177/15209156261507763
Vancouver
Weiskorn J, Auzanneau M, Burckhardt M, Blauensteiner N, Schierloh U, Mayer I, et al. The Varying Benefits of Two Years Automated Insulin Delivery System Use on Continuous Glucose Monitoring Profiles in Children and Adolescents with Type 1 Diabetes. Diabetes Technology & Therapeutics. 2026. doi:10.1177/15209156261507763
BibTeX
@article{Weiskorn2026,
author = {Weiskorn, Jantje and Auzanneau, Marie and Burckhardt, Marie-Anne and Blauensteiner, Nicole and Schierloh, Ulrike and Mayer, Isa and Schneidewind, Julia and Lanzinger, Stefanie},
title = {The Varying Benefits of Two Years Automated Insulin Delivery System Use on Continuous Glucose Monitoring Profiles in Children and Adolescents with Type 1 Diabetes},
journal = {Diabetes Technology & Therapeutics},
year = {2026},
doi = {10.1177/15209156261507763}
}