In vivo aging-induced surface roughness alterations of Invisalign ® and 3D-printed aligners
Journal of Orthodontics · 2023-12 · 50 (4) · 352-360
DOI: 10.1177/14653125221145948 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Orthodontics, online erschienen am 27.12.2022. Verfasst von Despina Koletsi und 5 weiteren Autor:innen. Beteiligte Einrichtung: Clinic of Orthodontics and Pediatric Dentistry, Center of Dental Medicine, University of Zurich, Zurich, Switzerland. Hauptthemen laut MEDLINE: Orthodontic Appliances, Removable. Die Arbeit wurde bisher 48-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objective
To assess the surface roughness of in-house 3D-printed orthodontic aligners compared with Invisalign ® appliances, both retrieved as well as in the ‘as-received’ control status.
Design
An in vitro study following intra-oral material aging. Setting and participants: Twelve clinically used Invisalign ® appliances and the same number of 3D-printed aligners, without involvement of attachments, were obtained from a respective number of patients. A similar number of ‘as-received’ aligners, of each material, were used as control (CON) groups.
Method
Four groups of materials were examined: A = Invisalign ® CON; B = Invisalign ® used; C = 3D-printed CON; and D = 3D-printed used. Optical profilometry was employed to examine the following surface roughness parameters: amplitude parameters Sa, Sq and Sz and functional parameters Sc and Sv. Descriptive statistics and quantile regression modeling were conducted, and the level of statistical significance was set at α = 0.05.
Results
Intra-oral exposure of 3D-printed aligners was significantly associated with increase in all tested parameters ( P < 0.001 at all occasions). Significant differences were detected in the retrieved 3D-printed aligners compared with Invisalign ® retrieved, with the exception of Sz. The respective effect sizes (median differences) were as follows: Sa: 169 nm, 95% confidence interval [CI] = 89–248, P < 0.001; Sq: 315 nm, 95% CI = 152–477, P < 0.001; Sc: 233 nm 3 /nm 2 , 95% CI = 131–335, P < 0.001; and Sv: 43 nm 3 /nm 2 , 95% CI = 17–68, P = 0.002.
Conclusion
Within the limitations of this study, we concluded that surface roughness differences existed between 3D-printed aligners and Invisalign ® in the retrieved status, as well as between the control and retrieved 3D-printed groups.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Clinic of Orthodontics and Pediatric Dentistry, Center of Dental Medicine, University of Zurich, Zurich, Switzerland.
- Department of Dentistry, School of Medicine, European University Cyprus, Nicosia, Cyprus.
- Private practice, Limassol, Cyprus.
- Department of Biomaterials, School of Dentistry, National and Kapodistrian University of Athens, Athens, Greece.
ORCID-Profile: Despina Koletsi, Athanasios E Athanasiou, Spiros Zinelis, Theodore Eliades
Zitieren
APA
Koletsi, D., Panayi, N., Laspos, C., Athanasiou, A. E., Zinelis, S., & Eliades, T. (2023). In vivo aging-induced surface roughness alterations of Invisalign ® and 3D-printed aligners. Journal of Orthodontics, 50(4), 352-360. https://doi.org/10.1177/14653125221145948
Vancouver
Koletsi D, Panayi N, Laspos C, Athanasiou AE, Zinelis S, Eliades T. In vivo aging-induced surface roughness alterations of Invisalign ® and 3D-printed aligners. Journal of Orthodontics. 2023;50(4):352-360. doi:10.1177/14653125221145948
BibTeX
@article{Koletsi2023,
author = {Koletsi, Despina and Panayi, Nearchos and Laspos, Christodoulos and Athanasiou, Athanasios E and Zinelis, Spiros and Eliades, Theodore},
title = {In vivo aging-induced surface roughness alterations of Invisalign ® and 3D-printed aligners},
journal = {Journal of Orthodontics},
year = {2023},
volume = {50},
number = {4},
pages = {352--360},
doi = {10.1177/14653125221145948}
}