EUVIMEDEUVIMED entdecken
Crossref

Ultrasound measurement of skin-to-septum distance in totally implantable venous access ports: Considerations for Huber needle length

Mari Abe iD, Takatoshi Kubo iD, Hikaru Haruki iD, Hironori Ishigami iD, Kazunaga Ishigaki, Ayako Suzuki, Gojiro Nakagami iD

The Journal of Vascular Access · 2026-11 · 27 (6) · 2797-2803

DOI: 10.1177/11297298261472621 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Journal of Vascular Access, online erschienen am 18.08.2026. Verfasst von Mari Abe und 6 weiteren Autor:innen. Beteiligte Einrichtung: Research Institute for Next-Generation Nursing Education, Tokyo, Japan. Die Arbeit nennt 12 Literaturangaben. Gefördert von: Terumo Foundation for Life Sciences and Arts.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Objective criteria for selecting Huber needle length for totally implantable venous access ports (TIVAPs) are limited. This study primarily described skin-to-septum distance using ultrasonography and its correlation with body habitus.

Methods

Thirty-three patients with TIVAPs for chemotherapy were included. Skin-to-septum distance was measured by ultrasonography in two positions: supine with arms at the sides and with shoulder flexion. Pearson’s correlation coefficient assessed correlation between body mass index (BMI) and septal depth. Excess needle length beyond the port base was calculated.

Results

Mean skin-to-septum distance was 1.7 ± 0.4 mm (arms at sides) and 1.8 ± 0.5 mm (shoulder flexion). A 16-mm Huber needle was used in 97.0%. The median excess needle length was 2.4 mm. BMI showed weak correlation with septal depth ( r = 0.33, p = 0.062).

Conclusions

Septal depth was shallow, and BMI alone was not a reliable predictor. Ultrasound may support Huber needle selection.

Schlagwörter

Stichwörter der Autor:innen

Huber needleTotally implantable venous access port (TIVAP)needle length selectionskin-to-septum distanceultrasound measurement

Autor:innen und Institutionen

  • Research Institute for Next-Generation Nursing Education, Tokyo, Japan.
  • Department of Nursing, The University of Tokyo Hospital, Tokyo, Japan.
  • Department of Gerontological Nursing/Wound Care Management, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan.
  • UTokyo Nursing Dissemination and Implementation Science Institute, Tokyo, Japan.
  • Department of Radiology, The University of Tokyo Hospital, Tokyo, Japan.
  • Department of Chemotherapy, The University of Tokyo Hospital, Tokyo, Japan.

ORCID-Profile: Mari Abe, Takatoshi Kubo, Hikaru Haruki, Hironori Ishigami, Gojiro Nakagami

Förderung

  • Terumo Foundation for Life Sciences and Arts

Zitieren

APA

Abe, M., Kubo, T., Haruki, H., Ishigami, H., Ishigaki, K., Suzuki, A., & Nakagami, G. (2026). Ultrasound measurement of skin-to-septum distance in totally implantable venous access ports: Considerations for Huber needle length. The Journal of Vascular Access, 27(6), 2797-2803. https://doi.org/10.1177/11297298261472621

Vancouver

Abe M, Kubo T, Haruki H, Ishigami H, Ishigaki K, Suzuki A, et al. Ultrasound measurement of skin-to-septum distance in totally implantable venous access ports: Considerations for Huber needle length. The Journal of Vascular Access. 2026;27(6):2797-2803. doi:10.1177/11297298261472621

BibTeX
@article{Abe2026,
  author = {Abe, Mari and Kubo, Takatoshi and Haruki, Hikaru and Ishigami, Hironori and Ishigaki, Kazunaga and Suzuki, Ayako and Nakagami, Gojiro},
  title = {Ultrasound measurement of skin-to-septum distance in totally implantable venous access ports: Considerations for Huber needle length},
  journal = {The Journal of Vascular Access},
  year = {2026},
  volume = {27},
  number = {6},
  pages = {2797--2803},
  doi = {10.1177/11297298261472621}
}
Kostenlos starten →