EUVIMEDEUVIMED entdecken
Fallbericht

Bedside removal of a knotted peripherally inserted central catheter using a Seldinger guidewire: Case report

Christian Ramacciani Isemann iD, Debora Verdi, Michela Passeri, Sabrina Signorini, Marco Matassa, Marco Fiorenzi

The Journal of Vascular Access · 2026

DOI: 10.1177/11297298251409466 ↗

Kurz gefasst

Fallbericht in der Zeitschrift The Journal of Vascular Access, 2026. Verfasst von Christian Ramacciani Isemann und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nursing and Midwifery, Azienda USL Toscana sud est, Arezzo, Italy. Hauptthemen laut MEDLINE: Device Removal, Catheterization, Central Venous, Catheters, Indwelling, Central Venous Catheters. Die Arbeit nennt 10 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Peripherally inserted central catheters (PICCs) rarely develop true knots, but when they do, forceful removal can precipitate catheter fracture and embolization. Bedside strategies that restore internal support before extraction may prevent escalation to invasive retrieval. An older adult female requiring prolonged intravenous antibiotics underwent ultrasound-guided right basilic PICC insertion (4 Fr, single lumen, polyurethane, power-injectable) using a standard Seldinger technique. After advancement through a peel-away sheath, flushing and aspiration were unsuccessful and intracavitary ECG did not show the expected P-wave augmentation. Gentle traction met immediate resistance. A 0.018-inch nitinol, straight-tip guidewire was reintroduced into the catheter to restore column strength; with controlled, gentle rotational traction the PICC was removed intact. Inspection showed a single loose knot approximately 2 cm proximal to the distal tip. A new PICC was placed contralaterally during the same session. No complications occurred and the patient was discharged without device-related sequelae. This case illustrates a pragmatic bedside maneuver for difficult PICC removal: pause traction, consider mechanical causes such as looping or knotting, reintroduce a guidewire to increase axial rigidity, and attempt controlled extraction while preparing an escalation plan. The approach may obviate fluoroscopic snare retrieval or surgical cut-down when there is no evidence of adherence, vascular injury, or partial fracture. When unexpected resistance is encountered during PICC removal, avoid forceful traction. Guidewire reintroduction can permit safe extraction of a knotted catheter and should be part of a stepwise troubleshooting algorithm.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Device RemovalCatheterization, Central VenousCatheters, IndwellingCentral Venous CathetersCatheterization, PeripheralHumansFemaleTreatment OutcomeEquipment DesignUltrasonography, InterventionalEquipment FailureAnti-Bacterial Agents

Stichwörter der Autor:innen

Catheter obstructionCatheterizationPeripherally inserted central cathetersUltrasonographyVascular access devicesinterventionalperipheral

Substanzen

Anti-Bacterial Agents

Autor:innen und Institutionen

  • Department of Nursing and Midwifery, Azienda USL Toscana sud est, Arezzo, Italy.
  • University of Siena, Siena, Italy.
  • Department of Emergency and Critical Care, Azienda USL Toscana sud est, Arezzo, Italy.

ORCID-Profile: Christian Ramacciani Isemann

Zitieren

APA

Ramacciani Isemann, C., Verdi, D., Passeri, M., Signorini, S., Matassa, M., & Fiorenzi, M. (2026). Bedside removal of a knotted peripherally inserted central catheter using a Seldinger guidewire: Case report. The Journal of Vascular Access. https://doi.org/10.1177/11297298251409466

Vancouver

Ramacciani Isemann C, Verdi D, Passeri M, Signorini S, Matassa M, Fiorenzi M. Bedside removal of a knotted peripherally inserted central catheter using a Seldinger guidewire: Case report. The Journal of Vascular Access. 2026. doi:10.1177/11297298251409466

BibTeX
@article{RamaccianiIsemann2026,
  author = {Ramacciani Isemann, Christian and Verdi, Debora and Passeri, Michela and Signorini, Sabrina and Matassa, Marco and Fiorenzi, Marco},
  title = {Bedside removal of a knotted peripherally inserted central catheter using a Seldinger guidewire: Case report},
  journal = {The Journal of Vascular Access},
  year = {2026},
  doi = {10.1177/11297298251409466}
}
Kostenlos starten →