A retrospective study of subcutaneous anchor securement systems in oncology patients
The Journal of Vascular Access · 2024
DOI: 10.1177/11297298231190416 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift The Journal of Vascular Access, 2024. Verfasst von Michelle L Hawes und 2 weiteren Autor:innen. Beteiligte Einrichtung: Data to Wisdom Research Consulting, Bloomington, IN, USA. Hauptthemen laut MEDLINE: Catheters, Indwelling, Central Venous Catheters, Catheterization, Central Venous, Catheterization, Peripheral. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Introduction
Maintaining optimal central venous catheter tip position requires reliable catheter securement. A vital decision about the choice of engineered securement device is often made by what is conveniently available in the insertion kit or default clinical routine. The importance of continuous securement for oncology patients prompted the need for an evaluation of securement options currently available. This study aimed to assess the effectiveness of two engineered securement devices to assist the oncology patient in reaching the end of their catheter need.
Methods
A retrospective study was conducted to assess patients’ ability to finish their therapy with one peripherally inserted central catheter. Implant and explant data for adult oncology patients was evaluated spanning 2007–2021. All patients received a PICC with either an adhesive securement device or a subcutaneous anchor securement system.
Results
Partial or complete dislodgement causing the unplanned removal of the PICC occurred at 12% for ASD and 0.4% for SASS ( p < 0.0001). The probability of reaching the end of need with one PICC, regardless of the reason for premature removal, at 2 years for patients with an adhesive securement device was 68% ( n = 944). For patients with a subcutaneous anchored securement device, it was over 95% ( n = 8313). The difference in the probability of reaching the end of the need with one PICC between the two securement devices was calculated at ( p < 0.0001).
Conclusion
With over 9200 patients and more than a million catheter days, the results of this retrospective study demonstrate the SASS’s superiority in assisting the patient to reach the end of need with a single PICC.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Data to Wisdom Research Consulting, Bloomington, IN, USA.
- Clatterbridge Cancer Center Liverpool, Liverpool.
- SigmaStats Consulting, LLC., Charleston, SC, USA.
ORCID-Profile: Michelle L Hawes
Zitieren
APA
Hawes, M. L., McCormick, C. A., & Gilbert, G. E. (2024). A retrospective study of subcutaneous anchor securement systems in oncology patients. The Journal of Vascular Access. https://doi.org/10.1177/11297298231190416
Vancouver
Hawes ML, McCormick CA, Gilbert GE. A retrospective study of subcutaneous anchor securement systems in oncology patients. The Journal of Vascular Access. 2024. doi:10.1177/11297298231190416
BibTeX
@article{Hawes2024,
author = {Hawes, Michelle L and McCormick, Carol A and Gilbert, Gregory E},
title = {A retrospective study of subcutaneous anchor securement systems in oncology patients},
journal = {The Journal of Vascular Access},
year = {2024},
doi = {10.1177/11297298231190416}
}
EUVIMED