Intracavitary EKG: A non-inferiority, prospective observational study comparing two methods, three devices
The Journal of Vascular Access · 2025
DOI: 10.1177/11297298251395136 ↗Kurz gefasst
Beobachtungsstudie in der Zeitschrift The Journal of Vascular Access, 2025. Verfasst von Federica Cirigliano und 15 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical and Clinical Sciences, Università degli Studi di Milano and Internal Medicine, Luigi Sacco Hospital, ASST-FBF-Sacco, Milan, …. Hauptthemen laut MEDLINE: Catheterization, Central Venous, Electrocardiography, Catheterization, Peripheral, Heart Rate. Die Arbeit nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: ATL Milano, Assago, Italy.
Abstract (Originalsprache)
Introduction
Accurate placement of central venous catheter tips is crucial for preventing complications, such as arrhythmias and vascular injuries. Intracavitary electrocardiography (I-EKG) has emerged as a reliable alternative to chest X-rays for catheter tip positioning, offering real-time results without radiation exposure. This study evaluates the quality of I-EKG tracings using two methods: the saline column method with the alligator clip and spindle system (Deltamed, Viadana, Italy) and the pure saline column method using the “red” connector cable (Vygon Card ® Vygon, Ecouen, France) and the “yellow” connector cable (Fast Tip ® ATL Milano, Assago, Italy).
Methods
Twelve patients undergoing a peripherally inserted central catheter (PICC) placement were enrolled in this prospective study at Luigi Sacco Hospital, Milan. I-EKG tracings were recorded using all three devices, yielding a total of 36 tracings. Objective measurements of P, QR, and RS wave amplitudes were taken over five heartbeats per tracing, while twelve experienced operators performed subjective evaluations using a 5-point scale. Statistical analyses compared the three techniques.
Results
There were no significant differences in tracing quality among the methods. Objective analyses of wave amplitude ratios confirmed equivalent performance across approaches. Subjective evaluations yielded comparable scores for the “alligator-spindle,” “red,” and “yellow” connector techniques, with no statistically significant variation.
Conclusions
With the limits of the small sample size, the pure saline column methods (“red” and “yellow” connectors) present similar EKG trace quality while offering procedural simplicity and enhanced safety. By eliminating the need for a spindle, these may reduce the risk of arrhythmia and avoid possible effects of guidewire contamination. These findings support the broader adoption of pure saline column techniques to optimize catheter placement procedures.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical and Clinical Sciences, Università degli Studi di Milano and Internal Medicine, Luigi Sacco Hospital, ASST-FBF-Sacco, Milan, Italy.
- Department of Infectious Diseases, "Luigi Sacco" Hospital, Milan, Italy.
- Università degli Studi di Milano and General Medicine II, San Paolo Hospital, ASST Santi Paolo e Carlo, Milan, Italy.
- Department of Internal Medicine, Division "Medicina Interna-Alta Complessità" Niguarda Hospital, ASST Grande Ospedale Metropolitano Niguarda, Milan, Italy.
- Division of Medicine and Liver/Gastroenterology Unit at San Paolo Hospital, ASST Santi Paolo e Carlo, Milan, Italy.
ORCID-Profile: Antonio Gidaro
Förderung
- ATL Milano, Assago, Italy
Zitieren
APA
Cirigliano, F., Taino, A., Bartoli, A., Busatto, N., Calloni, M., Casella, F., Donadoni, M., Foschi, A., La Cava, L., Masseroli, M. M., Perego, M., Salvi, E., Samartin, F., Zappa, P., Cogliati, C., & Gidaro, A. (2025). Intracavitary EKG: A non-inferiority, prospective observational study comparing two methods, three devices. The Journal of Vascular Access. https://doi.org/10.1177/11297298251395136
Vancouver
Cirigliano F, Taino A, Bartoli A, Busatto N, Calloni M, Casella F, et al. Intracavitary EKG: A non-inferiority, prospective observational study comparing two methods, three devices. The Journal of Vascular Access. 2025. doi:10.1177/11297298251395136
BibTeX
@article{Cirigliano2025,
author = {Cirigliano, Federica and Taino, Alba and Bartoli, Arianna and Busatto, Nicola and Calloni, Maria and Casella, Francesco and Donadoni, Mattia and Foschi, Antonella and La Cava, Leyla and Masseroli, Matteo Maria and Perego, Martina and Salvi, Emanuele and Samartin, Federica and Zappa, Paolo and Cogliati, Chiara and Gidaro, Antonio},
title = {Intracavitary EKG: A non-inferiority, prospective observational study comparing two methods, three devices},
journal = {The Journal of Vascular Access},
year = {2025},
doi = {10.1177/11297298251395136}
}