Hemodialysis access flow measurement: Comparison of ultrasound dilution and ultrafiltration method on NIKKISO DBB-EXA™ dialysis machine

Georgios Kosmadakis iD, Ioana Enache, Claudine Gueret, Abraham Haskour, Aura Necoara, Clémence Deville, Julien Baudenon, Nicole Rance

The Journal of Vascular Access · 2024

DOI: 10.1177/11297298231173284 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift The Journal of Vascular Access, 2024. Verfasst von Georgios Kosmadakis und 7 weiteren Autor:innen. Beteiligte Einrichtung: AURA Santé, Clermont-Ferrand, France. Hauptthemen laut MEDLINE: Arteriovenous Shunt, Surgical, Predictive Value of Tests, Renal Dialysis, Regional Blood Flow. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 25 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The methods of estimating vascular access (VA) flow rates are usually based on the indicator dilution theory by measuring recirculation during dialysis sessions.

Methods

This is an observational study comparing the VA flow rates measured by NIKKISO DBB-EXA™ and Transonic®. Sixty-five patients (38 M/27 F, mean age 72 ± 10 years) participated in the study. We measured the VA flow rates during dialysis twice with each method and repeated the procedure 7 days later.

Results

In 130 double measurements for each method on the same day, mean flow with Transonic® was 1413±715 ml/min and with DBB-EXA™ 1297 ± 664 ml/min. In Bland-Altman analysis, the mean difference between the two methods was 159 ± 211 ml/min (limits of agreement: −274 and 572 ml/min). Eighty-one out of the 130 DBB-EXA™ measurements were within 25% of the Transonic® measurements (62% accuracy). Regarding reproducibility of each method on different days, mean difference in the Bland-Altman analysis was 29 ± 620 ml/min (limits of agreement: −1186 and 1244 ml/min) for the Transonic® measurements and 132 ± 625 ml/min (limits of agreement: −1092 and 1356 ml/min) for the DBB-EXA™ measurements. The measurements on two different days were within 25% of each other for 52 of the 65 patients (80%) with the Transonic® method, and for 35 of the 65 patients (54%) with the DBB-EXA™ method.

Conclusions

In conclusion, the DBB-EXA™ method underestimates VA flow rates compared to the Transonic® technique, resulting in a limited accuracy of 62%. There was poor reproducibility for both methods in different day measurements with better performance of the Transonic® technique. The DBB-EXA™ method could be used as a simple tool for a rough estimate of VA flow rates but cannot replace the Transonic® reference method.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arteriovenous Shunt, SurgicalPredictive Value of TestsRenal DialysisRegional Blood FlowEquipment DesignHumansAgedFemaleMaleReproducibility of ResultsBlood Flow VelocityMiddle AgedAged, 80 and overTime FactorsIndicator Dilution TechniquesHemodiafiltrationBlood Vessel Prosthesis ImplantationTreatment Outcome

Stichwörter der Autor:innen

Vascular access flow ratearteriovenous fistula flow rateultrasound dilutionvascular access recirculation

Autor:innen und Institutionen

  • AURA Santé, Clermont-Ferrand, France.

ORCID-Profile: Georgios Kosmadakis

Zitieren

APA

Kosmadakis, G., Enache, I., Gueret, C., Haskour, A., Necoara, A., Deville, C., Baudenon, J., & Rance, N. (2024). Hemodialysis access flow measurement: Comparison of ultrasound dilution and ultrafiltration method on NIKKISO DBB-EXA™ dialysis machine. The Journal of Vascular Access. https://doi.org/10.1177/11297298231173284

Vancouver

Kosmadakis G, Enache I, Gueret C, Haskour A, Necoara A, Deville C, et al. Hemodialysis access flow measurement: Comparison of ultrasound dilution and ultrafiltration method on NIKKISO DBB-EXA™ dialysis machine. The Journal of Vascular Access. 2024. doi:10.1177/11297298231173284

BibTeX
@article{Kosmadakis2024,
  author = {Kosmadakis, Georgios and Enache, Ioana and Gueret, Claudine and Haskour, Abraham and Necoara, Aura and Deville, Clémence and Baudenon, Julien and Rance, Nicole},
  title = {Hemodialysis access flow measurement: Comparison of ultrasound dilution and ultrafiltration method on NIKKISO DBB-EXA™ dialysis machine},
  journal = {The Journal of Vascular Access},
  year = {2024},
  doi = {10.1177/11297298231173284}
}
Kostenlos starten →