Quantification of trunnion damage in a series of intact total hip arthroplasty femoral stems previously identified to be at risk of catastrophic failure
HIP International · 2024
DOI: 10.1177/11207000231199941 ↗Kurz gefasst
Fachartikel in der Zeitschrift HIP International, 2024. Verfasst von Brent A Lanting und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Surgery, London Health Sciences Centre - University Hospital, Division of Orthopaedic Surgery, ON, Canada. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Hip, Hip Prosthesis, Prosthesis Failure, Prosthesis Design. Die Arbeit nennt 35 Literaturangaben.
Abstract (Originalsprache)
Background
Corrosion at the head-neck junction of femoral stems is a rare complication of total hip arthroplasty (THA) with manifestations ranging from subclinical wear to failure. Prior studies have identified a single femoral component design with an increased propensity for catastrophic trunnion failure. The purpose of the present study was to quantify trunnion damage of this femoral component retrieved from patients undergoing revision THA for non-trunnionosis indications.
Methods
24 femoral components from a single manufacturer were identified for study inclusion. Each prosthesis underwent stereomicroscopic inspection. Corrosion and fretting scores were assigned per the Goldberg criteria to quadrants of the trunnion. Material loss was calculated based on cone angles across trunnion quadrants. This was carried out using a coordinate measuring machine that digitised each trunnion surface. Stems were compared to a series of femoral stems with the same trunnion design.
Results
20 of the 24 (83%) trunnions demonstrated corrosion, all 24 trunnions demonstrated fretting. Corrosion scores did not statistically differ with respect to trunnion zone ( p = 0.53), while fretting scores were higher in the inferior compared to the superior zones ( p < 0.001). There was no significant difference in cone angles assessing material loss between stems ( p = 0.25).
Conclusions
Evidence of trunnion damage was observed in each stem retrieved for non-trunnionosis revision. Fretting occurred more frequently about the inferior quadrants. However, digitised trunnion shapes were similar between compared stems exhibiting no material loss. Therefore, it is possible that previous reports of trunnion failures for this implant are not a systemic issue, and that further investigation is required.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Surgery, London Health Sciences Centre - University Hospital, Division of Orthopaedic Surgery, ON, Canada.
- Surgical Innovation Program, Lawson Health Research Institute, London, ON, Canada.
ORCID-Profile: Olawale A Sogbein, Nirmit Shah
Zitieren
APA
Lanting, B. A., Sogbein, O. A., MacDonald, S. J., Shah, N., Kok, T. L., Willing, R., & Teeter, M. G. (2024). Quantification of trunnion damage in a series of intact total hip arthroplasty femoral stems previously identified to be at risk of catastrophic failure. HIP International. https://doi.org/10.1177/11207000231199941
Vancouver
Lanting BA, Sogbein OA, MacDonald SJ, Shah N, Kok TL, Willing R, et al. Quantification of trunnion damage in a series of intact total hip arthroplasty femoral stems previously identified to be at risk of catastrophic failure. HIP International. 2024. doi:10.1177/11207000231199941
BibTeX
@article{Lanting2024,
author = {Lanting, Brent A and Sogbein, Olawale A and MacDonald, Steven J and Shah, Nirmit and Kok, Tea-Lyn and Willing, Ryan and Teeter, Matthew G},
title = {Quantification of trunnion damage in a series of intact total hip arthroplasty femoral stems previously identified to be at risk of catastrophic failure},
journal = {HIP International},
year = {2024},
doi = {10.1177/11207000231199941}
}
EUVIMED