EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Corrosion in Modular Revision Hip Stem Tapers – A Retrieval Analysis

Therese Bormann iD, Haolan Yan, Sebastian Jaeger, Mareike Schonhoff, J Philippe Kretzer

Journal of Orthopaedic Research · 2026

DOI: 10.1002/jor.70078 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Orthopaedic Research, 2026. Verfasst von Therese Bormann und 4 weiteren Autor:innen. Beteiligte Einrichtung: Research Center of Biomechanics and Implant Technology, Department of Orthopaedics, Heidelberg University Hospital, Germany. Hauptthemen laut MEDLINE: Hip Prosthesis, Prosthesis Failure, Arthroplasty, Replacement, Hip. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In revision hip arthroplasty, modular stems enable intraoperative adjustment of the biomechanics of the hip to ensure a stable joint function even in complex anatomical cases. Modular stem junctions, however, carry the risk of junction degradation due to corrosive processes or even junction breakage. Fretting‐corrosion has been mentioned as precursor of junction breakage but has hardly been systematically assessed. To investigate relations between corrosion and fretting, and implant and patient specific parameters and connection strength, respectively, a collection of 53 retrieved modular hip stems of different implant systems was investigated. Corrosion and fretting at the stem‐neck taper connection were rated with a modified Goldberg score. Taper contamination was assessed with a similar scoring system. If the hip stems were still joint to the neck piece, the push‐out force to detach the parts was recorded as a measure for taper junction strength. A multivariate regression analysis revealed that corrosion and fretting were significantly affected by implantation time, taper contamination, body weight and implant system. Taper junction strength was not altered by corrosion or fretting but by taper contamination. The results indicate that implant geometry parameters are not related to the extent of corrosive degradation at the junction, but taper contamination significantly increased corrosion at taper surfaces. This underlines the importance of the cleanliness of the taper surfaces during hip stem assembly for a long‐term stability of the modular implant.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Hip ProsthesisProsthesis FailureArthroplasty, Replacement, HipHumansCorrosionMaleFemaleMiddle AgedAgedReoperationProsthesis DesignAged, 80 and overAdult

Stichwörter der Autor:innen

corrosion wearfretting wearfretting‐corrosionrevision surgerytaper failuretotal hip arthroplasty

Autor:innen und Institutionen

  • Research Center of Biomechanics and Implant Technology, Department of Orthopaedics, Heidelberg University Hospital, Germany.

ORCID-Profile: Therese Bormann

Zitieren

APA

Bormann, T., Yan, H., Jaeger, S., Schonhoff, M., & Philippe Kretzer, J. (2026). Corrosion in Modular Revision Hip Stem Tapers – A Retrieval Analysis. Journal of Orthopaedic Research. https://doi.org/10.1002/jor.70078

Vancouver

Bormann T, Yan H, Jaeger S, Schonhoff M, Philippe Kretzer J. Corrosion in Modular Revision Hip Stem Tapers – A Retrieval Analysis. Journal of Orthopaedic Research. 2026. doi:10.1002/jor.70078

BibTeX
@article{Bormann2026,
  author = {Bormann, Therese and Yan, Haolan and Jaeger, Sebastian and Schonhoff, Mareike and Philippe Kretzer, J},
  title = {Corrosion in Modular Revision Hip Stem Tapers – A Retrieval Analysis},
  journal = {Journal of Orthopaedic Research},
  year = {2026},
  doi = {10.1002/jor.70078}
}
Kostenlos starten →