EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

A convolutional neural network for high throughput screening of femoral stem taper corrosion

Anastasia M Codirenzi, Brent A Lanting, Matthew G Teeter iD

Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine · 2023

DOI: 10.1177/09544119231177834 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2023. Verfasst von Anastasia M Codirenzi und 2 weiteren Autor:innen. Beteiligte Einrichtung: School of Biomedical Engineering, Western University, London, ON, Canada. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Hip, Hip Prosthesis. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Natural Sciences and Engineering Research Council of Canada.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Corrosion at the modular head-neck taper interface of total and hemiarthroplasty hip implants (trunnionosis) is a cause of implant failure and clinical concern. The Goldberg corrosion scoring method is considered the gold standard for observing trunnionosis, but it is labor-intensive to perform. This limits the quantity of implants retrieval studies typically analyze. Machine learning, particularly convolutional neural networks, have been used in various medical imaging applications and corrosion detection applications to help reduce repetitive and tedious image identification tasks. 725 retrieved modular femoral stem arthroplasty devices had their trunnion imaged in four positions and scored by an observer. A convolutional neural network was designed and trained from scratch using the images. There were four classes, each representing one of the established Goldberg corrosion classes. The composition of the classes were as follows: class 1 ( n = 1228), class 2 ( n = 1225), class 3 ( n = 335), and class 4 ( n = 102). The convolutional neural network utilized a single convolutional layer and RGB coloring. The convolutional neural network was able to distinguish no and mild corrosion (classes 1 and 2) from moderate and severe corrosion (classes 3 and 4) with an accuracy of 98.32%, a class 1 and 2 sensitivity of 0.9881, a class 3 and 4 sensitivity of 0.9556 and an area under the curve of 0.9740. This convolutional neural network may be used as a screening tool to identify retrieved modular hip arthroplasty device trunnions for further study and the presence of moderate and severe corrosion with high reliability, reducing the burden on skilled observers.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arthroplasty, Replacement, HipHip ProsthesisHumansFemur HeadCorrosionHigh-Throughput Screening AssaysReproducibility of ResultsProsthesis FailureProsthesis DesignPolyethylene

Stichwörter der Autor:innen

Machine learningarthroplastycorrosiontrunnionosis

Substanzen

Polyethylene

Autor:innen und Institutionen

  • School of Biomedical Engineering, Western University, London, ON, Canada.
  • Division of Orthopaedic Surgery, Department of Surgery, Schulich School of Medicine & Dentistry, Western University, London, ON, Canada.
  • Department of Medical Biophysics, Schulich School of Medicine & Dentistry, Western University, London, ON, Canada.

ORCID-Profile: Matthew G Teeter

Förderung

  • Natural Sciences and Engineering Research Council of Canada

Zitieren

APA

Codirenzi, A. M., Lanting, B. A., & Teeter, M. G. (2023). A convolutional neural network for high throughput screening of femoral stem taper corrosion. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. https://doi.org/10.1177/09544119231177834

Vancouver

Codirenzi AM, Lanting BA, Teeter MG. A convolutional neural network for high throughput screening of femoral stem taper corrosion. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 2023. doi:10.1177/09544119231177834

BibTeX
@article{Codirenzi2023,
  author = {Codirenzi, Anastasia M and Lanting, Brent A and Teeter, Matthew G},
  title = {A convolutional neural network for high throughput screening of femoral stem taper corrosion},
  journal = {Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine},
  year = {2023},
  doi = {10.1177/09544119231177834}
}
Kostenlos starten →