EUVIMEDEUVIMED entdecken
Crossref

Rapid point-of-care test for diagnosis of peritonitis in peritoneal dialysis patients

Htay Htay iD, Jason Chon Jun Choo, Dorothy Hannah Huang, Mathini Jayaballa, David W Johnson iD, Riece Koniman iD, Elizabeth Ley Oei, Tan Chieh Suai, Sin Yan Wu, Marjorie Wai Yin Foo

Peritoneal Dialysis International: Journal of the International Society for Peritoneal Dialysis · 2024-11 · 44 (6) · 413-418

DOI: 10.1177/08968608241234728 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Peritoneal Dialysis International: Journal of the International Society for Peritoneal Dialysis, online erschienen am 07.03.2024. Verfasst von Htay Htay und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Renal Medicine, Singapore General Hospital, Singapore, Singapore. Hauptthemen laut MEDLINE: Peritonitis, Peritoneal Dialysis, Interleukin-6, Point-of-Care Testing. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 10 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Periplex ® is a rapid point-of-care test based on the detection of interleukin-6 (IL-6) or matrix metalloproteinase-8 (MMP-8) to diagnose peritonitis in peritoneal dialysis (PD) patients.

Methods

This single-centre study was conducted in Singapore General Hospital from 2019 to 2022. The study recruited PD patients suspected of having peritonitis. Periplex was performed at the presentation and recovery of peritonitis. Primary outcomes were sensitivity and specificity of Periplex at presentation. The positive and negative predictive values of tests were also performed.

Results

A total of 120 patients were included in the study. The mean age was 60.9 ± 14.9 years, 53% were male, 79% were Chinese and 47.5% had diabetes mellitus. Periplex was positive in all patients with peritonitis ( n = 114); sensitivity of 100%; 95% confidence interval (CI): 100–100%. Periplex was falsely positive in three patients with non-infective eosinophilic peritonitis, resulting in a low specificity of 50%; 95% CI: 41.1–59.0%. Periplex had a positive predictive value of 97.4% and a negative predictive value of 100%. During recovery from peritonitis, Periplex had high specificity (93.6%) and negative predictive value (98.7%) to indicate the resolution of infection. MMP-8 was more sensitive than IL-6 in detecting peritonitis. Periplex was positive in all patients with peritonitis regardless of the types of PD solutions used.

Conclusions

Periplex had high sensitivity, and positive and negative predictive values in the diagnosis of peritonitis can be considered as a screening tool for peritonitis. Given its high specificity and negative predictive value, it may also be used to document the resolution of peritonitis.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PeritonitisPeritoneal DialysisInterleukin-6Point-of-Care TestingHumansMaleFemaleMiddle AgedSensitivity and SpecificityMatrix Metalloproteinase 8AgedPredictive Value of TestsSingaporeKidney Failure, Chronic

Stichwörter der Autor:innen

Biomarkersinterleukin-6 (IL-6)matrix metalloproteinase-8 (MMP-8)peritoneal dialysisperitonitispoint-of-care systemspoint-of-care testing

Substanzen

Interleukin-6Matrix Metalloproteinase 8

Autor:innen und Institutionen

  • Department of Renal Medicine, Singapore General Hospital, Singapore, Singapore.
  • DUKE-NUS Medical School, Singapore, Singapore.
  • Department of Nephrology, Princess Alexandra Hospital, Brisbane, QLD, Australia.
  • Centre for Kidney Disease Research, University of Queensland, Brisbane, QLD, Australia.

ORCID-Profile: Htay Htay, David W Johnson, Riece Koniman

Zitieren

APA

Htay, H., Choo, J. C. J., Huang, D. H., Jayaballa, M., Johnson, D. W., Koniman, R., Oei, E. L., Suai, T. C., Wu, S. Y., & Foo, M. W. Y. (2024). Rapid point-of-care test for diagnosis of peritonitis in peritoneal dialysis patients. Peritoneal Dialysis International: Journal of the International Society for Peritoneal Dialysis, 44(6), 413-418. https://doi.org/10.1177/08968608241234728

Vancouver

Htay H, Choo JCJ, Huang DH, Jayaballa M, Johnson DW, Koniman R, et al. Rapid point-of-care test for diagnosis of peritonitis in peritoneal dialysis patients. Peritoneal Dialysis International: Journal of the International Society for Peritoneal Dialysis. 2024;44(6):413-418. doi:10.1177/08968608241234728

BibTeX
@article{Htay2024,
  author = {Htay, Htay and Choo, Jason Chon Jun and Huang, Dorothy Hannah and Jayaballa, Mathini and Johnson, David W and Koniman, Riece and Oei, Elizabeth Ley and Suai, Tan Chieh and Wu, Sin Yan and Foo, Marjorie Wai Yin},
  title = {Rapid point-of-care test for diagnosis of peritonitis in peritoneal dialysis patients},
  journal = {Peritoneal Dialysis International: Journal of the International Society for Peritoneal Dialysis},
  year = {2024},
  volume = {44},
  number = {6},
  pages = {413--418},
  doi = {10.1177/08968608241234728}
}
Kostenlos starten →