Efficacy of a new detector automatic exposure control in portable anteroposterior chest radiography: a retrospective examination of radiation dose optimization and image quality
Acta Radiologica · 2026
DOI: 10.1177/02841851251387448 ↗Kurz gefasst
Fachartikel in der Zeitschrift Acta Radiologica, 2026. Verfasst von Yoon Ki Cha und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology and Center for Imaging Science, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea. Hauptthemen laut MEDLINE: Radiation Dosage, Radiography, Thoracic, Radiation Exposure. Die Arbeit nennt 9 Literaturangaben. Gefördert von: Future Medicine 20*30 Project of the Samsung Medical Center, The Korea Health Technology R&D Project through the Korea Health Industry Development Institute (KHIDI), funded by the Ministry of Health Welfare, Republic of….
Abstract (Originalsprache)
Background
Automatic exposure control (AEC) devices are necessary to reduce the radiation dose and enhance image quality in radiography. However, AEC devices are not being used in portable X-ray systems due to technical issues.
Purpose
To evaluate the radiation dose and image quality of a new AEC for a wireless portable X-ray system compared to the standard fixed radiation dose of manual controls in chest radiography. Material and Methods This retrospective analysis evaluated the performance of a portable X-ray system equipped with software-based AEC on anteroposterior chest radiographs. A quantitative evaluation comparing the manual and AEC groups was conducted to ascertain the exposure index (EI) and dose-area product (DAP). We performed observer-based analysis of image quality for lowest exposure and average exposure images in the manual and AEC groups.
Results
Overall, 2093 examinations were conducted in 467 patients. AEC yielded a statistically significant reduction in the DAP (AEC: 1.64 dGycm 2 ; manual: 1.99 dGycm 2 ) and EI (AEC: 266; manual: 393) compared to manual control ( P
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology and Center for Imaging Science, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
- Department of Radiology, Chung-Ang University Hospital, Seoul, Republic of Korea.
- Health & Medical Equipment Business, Samsung Electronics Co., Ltd, Seoul, Korea.
ORCID-Profile: Yoon Ki Cha, Myung Jin Chung
Förderung
- Future Medicine 20*30 Project of the Samsung Medical Center
- The Korea Health Technology R&D Project through the Korea Health Industry Development Institute (KHIDI), funded by the Ministry of Health Welfare, Republic of…
Zitieren
APA
Cha, Y. K., Woo, J. H., Kim, D. H., Kim, S., Lee, D. H., & Chung, M. J. (2026). Efficacy of a new detector automatic exposure control in portable anteroposterior chest radiography: a retrospective examination of radiation dose optimization and image quality. Acta Radiologica. https://doi.org/10.1177/02841851251387448
Vancouver
Cha YK, Woo JH, Kim DH, Kim S, Lee DH, Chung MJ. Efficacy of a new detector automatic exposure control in portable anteroposterior chest radiography: a retrospective examination of radiation dose optimization and image quality. Acta Radiologica. 2026. doi:10.1177/02841851251387448
BibTeX
@article{Cha2026,
author = {Cha, Yoon Ki and Woo, Jung Han and Kim, Dong Hyuk and Kim, Sanguk and Lee, Dong Hun and Chung, Myung Jin},
title = {Efficacy of a new detector automatic exposure control in portable anteroposterior chest radiography: a retrospective examination of radiation dose optimization and image quality},
journal = {Acta Radiologica},
year = {2026},
doi = {10.1177/02841851251387448}
}