Impact of reconstruction techniques on low dose chest CT image quality: comparison of FBP, Clear View at Mzuzu Central Hospital, Malawi
Malawi Medical Journal · 2025
DOI: 10.4314/mmj.v36i5.7 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Malawi Medical Journal, 2025. Verfasst von Hui Zhang und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Imaging, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, PR China. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Radiography, Thoracic, Radiographic Image Interpretation, Computer-Assisted, Lung. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Natural Science Foundation of Shaanxi Province, Key Research and Development Programs of Shaanxi Province, the Integration of Basic and Clinical Science Project of School of Basic Medical Sciences, Xi'an Jiaotong University.
Abstract (Originalsprache)
ObjectiveTo investigate the impact of two reconstruction techniques, Filtered Back Projection (FBP) and Clear View (CV) iterative algorithm, on the image quality of low-dose thin-slice chest CT. MethodsA retrospective study of 42 patients undergoing low-dose chest CT at Mzuzu Central Hospital from Feb-Apr 2024 used automatic tube current modulation at 120 kV. Raw data were reconstructed with FBP, 20% CV, 40% CV, 60% CV, and 80% CV, with 1 mm slice thickness and 0.625 mm spacing. Image noise, Signal-to-Noise Ratio (SNR), and Contrast-to-Noise Ratio (CNR) were measured, and image quality was rated on a 5-point scale for lung and mediastinal windows. Qualitative and quantitative parameters of the two different reconstruction algorithms in the five groups were comparatively analyzed.Results (1) Objective evaluation showed noise decreased in lung parenchyma, aorta, and erector spinae muscle with increasing CV weight. Mean noise reductions in lung parenchyma were 23.34% and 27.69% in 60% CV and 80% CV (P < 0.05). Aorta noise decreased by 23.43%, 37.16%, and 46.18% in 40% CV, 60% CV, and 80% CV (P < 0.05, P < 0.001, P < 0.001). Erector spinae muscle noise decreased by 35.91% and 44.78% in 60% CV and 80% CV (P < 0.05, P < 0.001). SNR and CNR were higher in CV groups than FBP. Among them, the differences in SNR between the 60% CV and 80% CV groups and the FBP group were statistically significant (P < 0.05). (2) Subjective scores for all groups were > 3, meeting diagnostic standards, with 60% CV yielding the highest lung and mediastinal window image quality (P < 0.05).ConclusionCompared to FBP, CV iterative reconstruction reduces noise and improves chest CT image quality under low-dose conditions as the weight increases, with 60% CV showing optimal performance.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medical Imaging, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, PR China.
- Radiology Department, Mzuzu Central Hospital, Mzuzu, Malawi.
- Department of Pathology, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, PR China.
- Department of Respiratory and Critical Care Medicine, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, PR China.
- Department of Cardiovascular Medicine, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an, PR China.
Förderung
- Natural Science Foundation of Shaanxi Province
- Key Research and Development Programs of Shaanxi Province
- the Integration of Basic and Clinical Science Project of School of Basic Medical Sciences, Xi'an Jiaotong University
Zitieren
APA
Zhang, H., Kondowe, B., Zhang, J., Xie, X., Song, Q., Niu, G., & Shang, J. (2025). Impact of reconstruction techniques on low dose chest CT image quality: comparison of FBP, Clear View at Mzuzu Central Hospital, Malawi. Malawi Medical Journal. https://doi.org/10.4314/mmj.v36i5.7
Vancouver
Zhang H, Kondowe B, Zhang J, Xie X, Song Q, Niu G, et al. Impact of reconstruction techniques on low dose chest CT image quality: comparison of FBP, Clear View at Mzuzu Central Hospital, Malawi. Malawi Medical Journal. 2025. doi:10.4314/mmj.v36i5.7
BibTeX
@article{Zhang2025,
author = {Zhang, Hui and Kondowe, Blessed and Zhang, Jiaojiao and Xie, Xinming and Song, Qiang and Niu, Gang and Shang, Jin},
title = {Impact of reconstruction techniques on low dose chest CT image quality: comparison of FBP, Clear View at Mzuzu Central Hospital, Malawi},
journal = {Malawi Medical Journal},
year = {2025},
doi = {10.4314/mmj.v36i5.7}
}
EUVIMED