Fully automated measurement of noise, signal-to-noise ratio, and contrast-to-noise ratio on chest CT images: feasibility and efficiency

Bozhe Mei iD, Zhangman Ma, Wanyun Fu, Linyang He, Zhicheng Ma, Xiangyang Gong iD

Acta Radiologica · 2024

DOI: 10.1177/02841851241287315 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Acta Radiologica, 2024. Verfasst von Bozhe Mei und 5 weiteren Autor:innen. Beteiligte Einrichtung: Jinzhou Medical University Postgraduate Education Base (Zhejiang Provincial People's Hospital, People's Hospital of Hangzhou Medical College), Hangzh…. Hauptthemen laut MEDLINE: Signal-To-Noise Ratio, Feasibility Studies, Tomography, X-Ray Computed, Radiography, Thoracic. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 23 Literaturangaben. Gefördert von: Key Research and Development Project of Zhejiang Province of China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Rapid and accurate measurement of computed tomography (CT) image noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) is a clinical challenge.

Purpose

To explore the feasibility of intelligent measurement of chest CT image noise, SNR, and CNR. Material and Methods A total of 300 chest CT scans were included in the study, which was divided into research dataset, internal test dataset, and external test dataset. Based on the research dataset, automatically segment and measure the average CT values and standard deviation (SD) of CT values for background air and lung field under different thresholds to obtain noise, SNR, and CNR results. Using the results of manual measurements as the reference standard, we determine the optimal threshold with the highest consistency. Using internal and external test datasets, validate the consistency of automated measurements of noise, SNR, and CNR at the optimal CT threshold with reference standards.

Results

With background air set at −900 HU and lung field at −800 HU as thresholds, the automated measurements of noise, SNR, and CNR demonstrate the highest consistency with the reference standards. At the optimal threshold, the noise, SNR, and CNR measured automatically on both the internal (intraclass correlation coefficient [ICC] = 0.85–0.96) and external (ICC = 0.75–0.85) test datasets exhibit high consistency with their respective reference standards.

Conclusion

The method we explored can intelligently measure the noise, SNR, and CNR of chest CT images, exhibits high consistency with radiologists, and offers a novel tool for image quality evaluation and analysis.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Signal-To-Noise RatioFeasibility StudiesTomography, X-Ray ComputedRadiography, ThoracicHumansMaleFemaleMiddle AgedAgedAdultRadiographic Image Interpretation, Computer-AssistedAged, 80 and overYoung AdultLungReproducibility of Results

Stichwörter der Autor:innen

Chest computed tomographyautomatic measurementnoisequality control

Autor:innen und Institutionen

  • Jinzhou Medical University Postgraduate Education Base (Zhejiang Provincial People's Hospital, People's Hospital of Hangzhou Medical College), Hangzhou, Zhejiang, PR China.
  • The Second School of Clinical Medicine, Zhejiang Chinese Medical University, Hangzhou, PR China.
  • Hangzhou Jianpei Technology Company Ltd, Hangzhou, PR China.
  • College of Computer Science and Technology, Zhejiang Gongshang University, Hangzhou, Zhejiang, PR China.
  • Department of Radiology, Center for Rehabilitation Medicine, Zhejiang Provincial People's Hospital (Affiliated People's Hospital), Hangzhou Medical College, Hangzhou, Zhejiang, PR China.

ORCID-Profile: Bozhe Mei, Xiangyang Gong

Förderung

  • Key Research and Development Project of Zhejiang Province of China

Zitieren

APA

Mei, B., Ma, Z., Fu, W., He, L., Ma, Z., & Gong, X. (2024). Fully automated measurement of noise, signal-to-noise ratio, and contrast-to-noise ratio on chest CT images: feasibility and efficiency. Acta Radiologica. https://doi.org/10.1177/02841851241287315

Vancouver

Mei B, Ma Z, Fu W, He L, Ma Z, Gong X. Fully automated measurement of noise, signal-to-noise ratio, and contrast-to-noise ratio on chest CT images: feasibility and efficiency. Acta Radiologica. 2024. doi:10.1177/02841851241287315

BibTeX
@article{Mei2024,
  author = {Mei, Bozhe and Ma, Zhangman and Fu, Wanyun and He, Linyang and Ma, Zhicheng and Gong, Xiangyang},
  title = {Fully automated measurement of noise, signal-to-noise ratio, and contrast-to-noise ratio on chest CT images: feasibility and efficiency},
  journal = {Acta Radiologica},
  year = {2024},
  doi = {10.1177/02841851241287315}
}
Kostenlos starten →