Potential of photon-counting detector CT technology for contrast medium reduction in portal venous phase thoracoabdominal CT
European Radiology · 2025
DOI: 10.1007/s00330-025-11409-3 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Radiology, 2025. Verfasst von Daniel Popp und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic and Interventional Radiology, University Hospital Augsburg, Faculty of Medicine, University of Augsburg, Augsburg, Germany. Hauptthemen laut MEDLINE: Contrast Media, Tomography, X-Ray Computed, Portal Vein, Radiography, Abdominal. Die Arbeit wurde bisher 15-mal zitiert (Crossref) und nennt 48 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objectives
To compare image quality and iodine attenuation intra-individually in portal venous phase photon-counting detector CT (PCD-CT) scans using protocols with different contrast medium (CM) volume. Materials and methods A prospectively acquired patient cohort between 04/2021 and 11/2023 was retrospectively screened if patients had the following combination of portal venous phase thoracoabdominal CT scans: (a) PCD-CT with 120 mL CM volume (PCD-CT 120 mL ), (b) PCD-CT with 100 mL CM volume (PCD-CT 100 mL ), and (c) prior energy-integrating detector CT (EID-CT) with 120 mL CM volume. On PCD-CT, virtual monoenergetic image (VMI) reconstructions at 70 keV were applied for both groups as well as additional VMI at 60 keV for PCD‑CT 100 mL . Quantitative analyses including signal-to-noise (SNR) and contrast-to-noise ratios (CNR) and qualitative analyses were performed using a mixed linear effects model.
Results
The final study cohort comprised 49 patients (mean age 67 [31–86] years, 12 female). Comparison to EID-CT was available in 33 patients. In standard 70 keV VMI reconstructions, PCD-CT 100 mL was non-inferior to PCD-CT 120 mL as well as to EID-CT 120 mL for CNR in abdominal organs (all p > 0.050). The mixed linear effects model revealed significant differences between contrast volume groups for both contrast enhancement and image quality ratings. PCD-CT 100 mL/70 keV demonstrated the smallest deviation from optimal contrast enhancement (−0.306, p < 0.001).
Conclusion
In portal venous phase thoracoabdominal PCD-CT, a nearly 17% reduction in CM was achievable while maintaining subjective and objective image quality compared to prior higher CM volume PCD-CT scans within the same patients and still surpassing image quality of previous exams on an EID-CT system. Key Points Question How do image quality and iodine attenuation intra-individually compare in portal venous phase photon-counting detector CT (PCD-CT) scans using protocols with different contrast medium volume.
Findings
PCD-CT scans exhibit superior quantitative and qualitative image quality compared to energy-integrating detector-CT acquisitions and are not negatively affected by contrast volume reductions up to 17%. Clinical relevance This study provides further evidence that PCD-CT enables a considerable reduction in iodine dose for portal venous phase acquisition, benefiting both patients and healthcare system costs.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Diagnostic and Interventional Radiology, University Hospital Augsburg, Faculty of Medicine, University of Augsburg, Augsburg, Germany.
- Department of Radiology and Nuclear medicine, Kantonspital St. Gallen, St. Gallen, Switzerland.
- Institute for Radiology, DONAUISAR Hospital Deggendorf-Dingolfing-Landau, Deggendorf, Germany.
- Medical Faculty, Ludwig Maximilian University Munich, Munich, Germany.
- Department of Diagnostic and Interventional Radiology, University Hospital Augsburg, Faculty of Medicine, University of Augsburg, Augsburg, Germany. Thomas.Kroencke@uk-augsburg.de.
- Centre for Advanced Analytics and Predictive Sciences (CAAPS), University of Augsburg, Augsburg, Germany. Thomas.Kroencke@uk-augsburg.de.
Zitieren
APA
Popp, D., Siedlecki, M., Friedrich, L., Haerting, M., Scheurig-Muenkler, C., Schwarz, F., Kroencke, T., Bette, S., & Decker, J. A. (2025). Potential of photon-counting detector CT technology for contrast medium reduction in portal venous phase thoracoabdominal CT. European Radiology. https://doi.org/10.1007/s00330-025-11409-3
Vancouver
Popp D, Siedlecki M, Friedrich L, Haerting M, Scheurig-Muenkler C, Schwarz F, et al. Potential of photon-counting detector CT technology for contrast medium reduction in portal venous phase thoracoabdominal CT. European Radiology. 2025. doi:10.1007/s00330-025-11409-3
BibTeX
@article{Popp2025,
author = {Popp, Daniel and Siedlecki, Martin and Friedrich, Lena and Haerting, Mark and Scheurig-Muenkler, Christian and Schwarz, Florian and Kroencke, Thomas and Bette, Stefanie and Decker, Josua A},
title = {Potential of photon-counting detector CT technology for contrast medium reduction in portal venous phase thoracoabdominal CT},
journal = {European Radiology},
year = {2025},
doi = {10.1007/s00330-025-11409-3}
}
EUVIMED