Trade-off between the radiation parameters and image quality using iterative reconstruction techniques in head computed tomography: a phantom study

Elīza Lāce, Reza Mohammadian iD, Ainārs Āboltiņš, Dāvis Sosārs, Ilze Apine

Acta Radiologica · 2023

DOI: 10.1177/02841851231185347 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Acta Radiologica, 2023. Verfasst von Elīza Lāce und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Riga Stradin's University, Riga, Latvia. Hauptthemen laut MEDLINE: Radiographic Image Interpretation, Computer-Assisted, Tomography, X-Ray Computed. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 33 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Iterative reconstruction techniques (IRTs) are commonly used in computed tomography (CT) and help to reduce image noise.

Purpose

To determine the minimum radiation dose while preserving image quality in head CT using IRTs. Material and Methods The anthropomorphic phantom was used to scan nine head CT image series with varied radiation parameters. CT dose parameters, including volume CT dose index (CTDIvol [in mGy]) and dose length product (DLP [in mGy/cm]), were recorded for each scan series. Different noise levels (iDoseL1-6) were used in IRT reconstructions for soft and bone tissues. In total, 15 measurements were taken from five regions of interest (ROI) with an area of 10 mm 2 . The signal-to-noise ratio (SNR) and noise values obtained at different ROIs were compared among various reconstruction methods with repeated measures of statistical analysis .

Results

In the head CT scan, applying IRT iDoseL5 had the lowest noise and highest SNR for soft tissue ( P < 0.05), and increased iDose can decrease CT dose by 54.6% without compromising image quality. While for bone tissue reconstruction, no clear association was found between the level of iDose and noise. However, when CTDIvol is >20 mGy, iDoseL4 is slightly superior to other reconstruction methods ( P < 0.065).

Conclusion

Using IRTs in head CTs reduces radiation dose while maintaining image quality. IDoseL5 provided optimal balance for soft tissue.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiographic Image Interpretation, Computer-AssistedTomography, X-Ray ComputedHumansRadiation DosageHeadSignal-To-Noise RatioAlgorithms

Stichwörter der Autor:innen

Dose optimizationiDoseiterative reconstruction techniques

Autor:innen und Institutionen

  • Department of Radiology, Riga Stradin's University, Riga, Latvia.
  • Department of Radiology, Children's Clinical University Hospital, Riga, Latvia.

ORCID-Profile: Reza Mohammadian

Zitieren

APA

Lāce, E., Mohammadian, R., Āboltiņš, A., Sosārs, D., & Apine, I. (2023). Trade-off between the radiation parameters and image quality using iterative reconstruction techniques in head computed tomography: a phantom study. Acta Radiologica. https://doi.org/10.1177/02841851231185347

Vancouver

Lāce E, Mohammadian R, Āboltiņš A, Sosārs D, Apine I. Trade-off between the radiation parameters and image quality using iterative reconstruction techniques in head computed tomography: a phantom study. Acta Radiologica. 2023. doi:10.1177/02841851231185347

BibTeX
@article{Lce2023,
  author = {Lāce, Elīza and Mohammadian, Reza and Āboltiņš, Ainārs and Sosārs, Dāvis and Apine, Ilze},
  title = {Trade-off between the radiation parameters and image quality using iterative reconstruction techniques in head computed tomography: a phantom study},
  journal = {Acta Radiologica},
  year = {2023},
  doi = {10.1177/02841851231185347}
}
Kostenlos starten →