Reduction in Sample Recollection Number Following the Implementation of Automated Pre-Analytical Interference Detection in Coagulation Testing

Vitória Luiza Camargo Milczarski, Iuri Vicente Camargo Morkis, Camila da Silveira Mariot, Gabriel Giron Correa, Carine Ghem

Annals of Clinical Biochemistry: International Journal of Laboratory Medicine · 2026

Vollständiger Abstract

Worum geht es in dieser Arbeit?

Background: Pre-analytical interferents such as hemolysis, lipemia, and icterus are major sources of error in coagulation testing and frequently lead to sample recollection. Visual inspection, although widely used, is subjective and may fail to identify unsuitable samples. Automated pre-analytical interferent detection using hemolysis, icterus, and lipemia (HIL) indices has been introduced to improve reliability and laboratory efficiency. Objective: Evaluate the impact of automated pre-analytical interferent detection on the number of sample recollections due to hemolysis in routine coagulation tests. Methods: A retrospective study compared two one-year periods in a tertiary hospital laboratory. In 2022, samples were assessed by visual inspection using the STA-R® analyzer, while in 2023 automated inspection was implemented using the ACL Top® 750 system. Recollections due exclusively to hemolysis were analyzed for PT, aPTT, FV, fibrinogen, and D-dimer tests. Statistical analyses included binomial, Wald, and Student’s t tests, with p ≤ 0.005 considered significant. Results: A significant reduction in recollections was observed after implementation of automated inspection (0.49% vs. 0.35%; p < 0.001). PT showed a statistically significant decrease in recollections (p = 0.003), while other tests demonstrated non-significant reductions. Monthly average recollections decreased by 18.9% (p < 0.001). Annual costs related to recollections were reduced from US$ 638.49 to US$ 399.74, representing savings of US$ 238.75. A significant reduction was also observed in emergency department recollections. Conclusions: Automated pre-analytical interferent detection significantly reduced hemolysis-related recollections, costs, and subjectivity in coagulation testing, supporting its implementation as a valuable tool to improve laboratory workflow and result reliability.

Bibliografischer Nachweis

Publikationsdaten

Autor:innen
Vitória Luiza Camargo Milczarski, Iuri Vicente Camargo Morkis, Camila da Silveira Mariot, Gabriel Giron Correa, Carine Ghem
Quelle
Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
Publikation
2026-01-01
Band / Ausgabe
Nicht angegeben
Seiten
Nicht angegeben
ISSN / ISBN
0004-5632, 1758-1001
Zitationen
0 laut Crossref
Referenzen
0 hinterlegt

Zitieren

Zitierfähiger Nachweis

Vitória Luiza Camargo Milczarski, Iuri Vicente Camargo Morkis, Camila da Silveira Mariot, Gabriel Giron Correa, Carine Ghem (2026). Reduction in Sample Recollection Number Following the Implementation of Automated Pre-Analytical Interference Detection in Coagulation Testing. Annals of Clinical Biochemistry: International Journal of Laboratory Medicine. https://doi.org/10.1177/00045632261488734
RIS BibTeX CSL-JSON

Kontext

Themen, Förderung und Nutzung

Lizenzhinweise: Lizenz 1