Repeatability and Cross-Domain Agreement of Pyramidal-Sensor Ocular Wavefront Measurements in Myopic Eyes: A Registry Validation of OSIRIS
Translational Vision Science & Technology · 29.10.2026 · 15 (10) · 49
DOI: 10.1167/tvst.15.10.49 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Translational Vision Science & Technology, erschienen am 29.10.2026. Verfasst von Ivan Gabric und 2 weiteren Autor:innen. Beteiligte Einrichtung: Eye Clinic Svjetlost, Zagreb, Croatia. Hauptthemen laut MEDLINE: Aberrometry, Registries, Myopia, Refraction, Ocular. Die Arbeit nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
The purpose of this study was to quantify within-session repeatability of ocular wavefront metrics from a pyramidal wavefront-sensor aberrometer (OSIRIS; CSO), to define a pupil-support and zero-code gate for 6-mm analysis, and to assess agreement between OSIRIS whole-eye and MS-39 total-corneal wavefront.
Methods
Retrospective single-center registry analysis of myopic and myopic astigmatic eyes. Within-subject standard deviation (SD), repeatability coefficient (RC = 2.77 × SD), and intraclass correlation coefficient ICC (1.1) were computed across within-session scan clusters with cluster-bootstrap confidence intervals (CIs). Agreement used Bland-Altman limits and Lin concordance correlation coefficient (CCC); and astigmatism used Thibos power vectors.
Results
Across 10,416 sessions from 2533 eyes, single-scan repeatability was excellent: 4-mm higher-order aberration (HOA) root mean square (RMS), ICC = 0.987 (RC = 0.033 µm); pupil-supported 6-mm HOA RMS ICC = 0.992 (RC = 0.050 µm); and 6-mm spherical aberration ICC = 0.997. Preoperative eyes were the noisier stratum (6-mm HOA RC = 0.065 vs. 0.039 postoperatively). Diameters below 6 mm gave entirely zero-coded 6-mm exports in 96.8% of sessions. Omitting scan-selection worsened 6-mm HOA RC to 0.257 µm. Whole-eye and total-corneal values correlated but were not interchangeable (6-mm HOA RMS CCC = 0.623). Objective and manifest refraction were consistent (4-mm sphere CCC = 0.965), a rounding-limited consistency check.
Conclusions
In myopic eyes, OSIRIS shows excellent within-session repeatability under the described acquisition and quality-control (QC) workflow. Six-millimeter analysis requires pupil-diameter and zero-code gating. Whole-eye and total-corneal wavefront are not interchangeable. Translational Relevance: Within-session precision measured in routine high-volume practice, with a pupil-diameter and zero-code quality rule, defines the smallest same-session change distinguishable from measurement variability, informing wavefront-guided planning, lens selection, and post-refractive follow-up.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Eye Clinic Svjetlost, Zagreb, Croatia.
- University of Zagreb School of Medicine, Zagreb, Croatia.
Zitieren
APA
Gabric, I., Modric, I. S., & Grulovic, F. (2026). Repeatability and Cross-Domain Agreement of Pyramidal-Sensor Ocular Wavefront Measurements in Myopic Eyes: A Registry Validation of OSIRIS. Translational Vision Science & Technology, 15(10), 49. https://doi.org/10.1167/tvst.15.10.49
Vancouver
Gabric I, Modric IS, Grulovic F. Repeatability and Cross-Domain Agreement of Pyramidal-Sensor Ocular Wavefront Measurements in Myopic Eyes: A Registry Validation of OSIRIS. Translational Vision Science & Technology. 2026;15(10):49. doi:10.1167/tvst.15.10.49
BibTeX
@article{Gabric2026,
author = {Gabric, Ivan and Modric, Ivan Samuel and Grulovic, Filip},
title = {Repeatability and Cross-Domain Agreement of Pyramidal-Sensor Ocular Wavefront Measurements in Myopic Eyes: A Registry Validation of OSIRIS},
journal = {Translational Vision Science & Technology},
year = {2026},
volume = {15},
number = {10},
pages = {49},
doi = {10.1167/tvst.15.10.49}
}
EUVIMED