Enhancing Clinical Full-Field Electroretinograms by Reducing Powerline Noise Using Pre-Amplified Electrodes

Tony T C Man, Yolanda W Y Yip, Chi Pui Pang, Mårten Erik Brelén

Translational Vision Science & Technology · 2025

DOI: 10.1167/tvst.14.10.27 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Translational Vision Science & Technology, 2025. Verfasst von Tony T C Man und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Kowloon, Hong Kong. Hauptthemen laut MEDLINE: Electroretinography, Retina. Die Arbeit nennt 11 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This study aimed to evaluate the efficacy of pre-amplified electrodes in mitigating powerline noise interference in clinical full-field electroretinograms (ffERGs). Pre-amplified electrodes can reduce powerline noise by providing an alternate pathway for the dissipation of electromagnetically induced noise. Building upon our previous theoretical and animal research, this investigation sought to validate these findings in a clinical setting.

Methods

Twenty-nine normal-sighted subjects participated in the study. Standard ffERGs were obtained using both traditional passive electrodes and pre-amplified electrodes consecutively from the same eye. Signal metrics and repeatability were evaluated by comparing the amplitudes and peak times of a-wave and b-wave components. Powerline noise was analyzed using frequency spectrum analysis.

Results

The mean amplitudes and peak times of a-wave and b-wave components between the pre-amplified and passive electrodes showed no statistically significant differences (P > 0.05). Intrasubject variability was modestly reduced when using pre-amplified electrodes, as indicated by lower intra-subject standard deviation (P 2.25) in flash stimulation setups.

Conclusions

Whereas the intrasubject variability improvements in the normally sighted subjects were modest, the improvements in the signal to noise ratio suggest that effective mitigation of powerline noise are possible with pre-amplified electrodes in diseased subjects with reduced ERG amplitudes. Translational Relevance: A pre-amplified electrode can reliably and robustly reduce powerline noise and thus improve signal repeatability while preserving signal integrity in clinical ffERGs.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ElectroretinographyRetinaHumansMaleSignal-To-Noise RatioAdultFemaleElectrodesYoung AdultReproducibility of ResultsMiddle Aged

Autor:innen und Institutionen

  • Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Kowloon, Hong Kong.
  • Pao So Kok Macular Disease Treatment and Research Centre, Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Hong Kong.

Zitieren

APA

Man, T. T. C., Yip, Y. W. Y., Pang, C. P., & Brelén, M. E. (2025). Enhancing Clinical Full-Field Electroretinograms by Reducing Powerline Noise Using Pre-Amplified Electrodes. Translational Vision Science & Technology. https://doi.org/10.1167/tvst.14.10.27

Vancouver

Man TTC, Yip YWY, Pang CP, Brelén ME. Enhancing Clinical Full-Field Electroretinograms by Reducing Powerline Noise Using Pre-Amplified Electrodes. Translational Vision Science & Technology. 2025. doi:10.1167/tvst.14.10.27

BibTeX
@article{Man2025,
  author = {Man, Tony T C and Yip, Yolanda W Y and Pang, Chi Pui and Brelén, Mårten Erik},
  title = {Enhancing Clinical Full-Field Electroretinograms by Reducing Powerline Noise Using Pre-Amplified Electrodes},
  journal = {Translational Vision Science & Technology},
  year = {2025},
  doi = {10.1167/tvst.14.10.27}
}
Kostenlos starten →