Measuring differences in the ERG in myopia using the RET eval device with skin electrodes
Ophthalmic and Physiological Optics · 2025
DOI: 10.1111/opo.13460 ↗Kurz gefasst
Fachartikel in der Zeitschrift Ophthalmic and Physiological Optics, 2025. Verfasst von Victoria Stapley und 3 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Optometry and Vision Science, Biomedical Sciences Research Institute, Ulster University, Coleraine, UK. Hauptthemen laut MEDLINE: Myopia, Electroretinography, Refraction, Ocular, Dark Adaptation. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 84 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Department for the Economy.
Abstract (Originalsprache)
Introduction
Previous research suggests that the electroretinogram (ERG) is reduced and delayed in non‐pathological myopia. However, the invasive nature of the electrode and cumbersome equipment required has prevented the widescale uptake of ERG measures. This study investigated whether previously reported changes to the ERG response in myopia are also observable when measured using non‐invasive skin electrodes and a hand‐held ERG device.
Method
Monocular flash ERGs were measured using the RET eval ® device according to the ‘ISCEV 6 Step Dark First cd’ protocol in 46 participants with non‐pathological myopia (spherical equivalent refraction [SER] −0.50 to −11.25 D, median −3.75 D, median axial length [AL] 25.4 mm) and 47 non‐myopic controls (SER +2.00 to −0.25 D, median +1.00 D, median AL 23.6 mm). Measures were performed under pupil mydriasis with Sensor Strip skin electrodes.
Results
The median implicit time for all dark‐adapted (DA) components was longer among myopes. Following Holm‐Bonferroni correction, this difference reached statistical significance ( p < 0.05) for the DA 3.0 A‐wave, DA 10.0 A‐wave and B‐waves, and DA Oscillatory potentials 1 and 2. There were no significant differences between median light‐adapted (LA) implicit times nor response amplitudes between refractive groups. For all DA components, there was a significant, positive correlation between AL and implicit time (all p < 0.05).
Conclusions
The RET eval ®, used with skin electrodes, did not detect the reduction in ERG amplitude reported in myopic eyes using traditional ERG setups, potentially due to high inter‐subject variability and/or anatomical confounders associated with the use of a skin electrode. The RET eval ® with skin electrodes did detect subtle delays to DA implicit times previously reported in myopia, with a positive relationship observed between AL and implicit time for all DA components. In contrast, no significant differences were observed for LA implicit times, which may indicate underlying differences in the dark‐adaptation process and/or scotopic visual pathways in myopia.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Centre for Optometry and Vision Science, Biomedical Sciences Research Institute, Ulster University, Coleraine, UK.
- National Institute for Health and Care Research (NIHR) Biomedical Research Centre at Moorfields Eye Hospital NHS Foundation Trust and UCL Institute of Ophthalmology, London, UK, London, UK.
ORCID-Profile: Victoria Stapley, Kathryn Saunders
Förderung
- Department for the Economy
Zitieren
APA
Stapley, V., Anderson, R. S., Saunders, K., & Mulholland, P. J. (2025). Measuring differences in the ERG in myopia using the RET eval device with skin electrodes. Ophthalmic and Physiological Optics. https://doi.org/10.1111/opo.13460
Vancouver
Stapley V, Anderson RS, Saunders K, Mulholland PJ. Measuring differences in the ERG in myopia using the RET eval device with skin electrodes. Ophthalmic and Physiological Optics. 2025. doi:10.1111/opo.13460
BibTeX
@article{Stapley2025,
author = {Stapley, Victoria and Anderson, Roger S and Saunders, Kathryn and Mulholland, Pádraig J},
title = {Measuring differences in the ERG in myopia using the RET eval device with skin electrodes},
journal = {Ophthalmic and Physiological Optics},
year = {2025},
doi = {10.1111/opo.13460}
}
EUVIMED