Influence of Lens Power on IOL/Posterior Lens Capsule Interactions and IOL's PCO Potential
Investigative Ophthalmology & Visual Science · 2025
DOI: 10.1167/iovs.66.1.41 ↗At a glance
Journal article in Investigative Ophthalmology & Visual Science, 2025. Written by Amjad Chatila and 7 co-authors. Institution: Department of Bioengineering, University of Texas at Arlington, Arlington, Texas, United States. Major MEDLINE topics: Capsule Opacification, Lenses, Intraocular, Posterior Capsule of the Lens, Myopia. Cited 3 times so far (Crossref); 66 references. The full text is freely available (open access, CC BY-NC-ND 4.0). Funded by: NHLBI NIH HHS.
Abstract (original language)
Purpose
Severely myopic eyes have been associated with high posterior capsule opacification (PCO) incidence. Although it has been reported that myopic eyes have weaker or more delayed capsule adhesion than emmetropic eyes, it is unclear whether/how dioptric power and posterior curvature of IOLs affect IOLs' affinity for the posterior lens capsule (PLC) and their PCO potential.
Methods
To investigate this, acrylic foldable IOLs with increasing dioptric power of 6.0D (for high myopia), 20.0D, and 30.0D (for low/non-myopia) were tested on their binding affinity toward PLC and their ability to inhibit the proliferation and infiltration of lens epithelial cells (LECs) using an in vitro simulated human PLC (sPLC) model.
Results
We found that IOL power and posterior radius of curvature (PRC) had significant impacts on IOL/sPLC adhesion forces, which are in the following order: 20.0D ≈ 30.0D > 6.0D. Optical coherence tomography (OCT) images showed that loose binding between 6.0D IOLs and sPLC contributed to larger interface spaces and significantly greater LEC infiltration, proliferation, metabolic activity, and transdifferentiation compared to 20.0D and 30.0D IOLs. Statistical analyses showed that IOLs' PRC may have a substantial influence on IOL/sPLC physical interactions, LEC responses, and PCO incidence.
Conclusions
The overall results suggest that the high PRC of low-diopter (6.0D) IOLs reduces their binding affinity toward the PLC, facilitates LEC reactions, thus causes high PCO incidence in myopic eyes. These findings strongly support that a new design to increase IOL posterior surfaces' PLC affinity may reduce PCO potential and increase safety for myopic patients.
Subjects
MeSH terms (MEDLINE)
Authors and institutions
- Department of Bioengineering, University of Texas at Arlington, Arlington, Texas, United States.
Funding
- NHLBI NIH HHS
Cite
APA
Chatila, A., Ea, V., Izuagbe, S., Hoang, L. Q., Vaish, B., Co, C. M., Luong, T. M., & Tang, L. (2025). Influence of Lens Power on IOL/Posterior Lens Capsule Interactions and IOL's PCO Potential. Investigative Ophthalmology & Visual Science. https://doi.org/10.1167/iovs.66.1.41
Vancouver
Chatila A, Ea V, Izuagbe S, Hoang LQ, Vaish B, Co CM, et al. Influence of Lens Power on IOL/Posterior Lens Capsule Interactions and IOL's PCO Potential. Investigative Ophthalmology & Visual Science. 2025. doi:10.1167/iovs.66.1.41
BibTeX
@article{Chatila2025,
author = {Chatila, Amjad and Ea, Vicki and Izuagbe, Samira and Hoang, Le Quynh and Vaish, Bhavya and Co, Cynthia M and Luong, Trinh M and Tang, Liping},
title = {Influence of Lens Power on IOL/Posterior Lens Capsule Interactions and IOL's PCO Potential},
journal = {Investigative Ophthalmology & Visual Science},
year = {2025},
doi = {10.1167/iovs.66.1.41}
}
EUVIMED