Abstract FR426: Loudness Does Not Predict Inter-Rater Blood Pressure Agreement in Digitally Captured Auscultatory Recordings
Hypertension · 2026-10 · 83 (Suppl_1)
DOI: 10.1161/hyp.83.suppl_1.fr426 ↗Kurz gefasst
Fachartikel in der Zeitschrift Hypertension, 2026-10. Verfasst von Tuhin Sinha und 1 weiteren Autor:innen.
Abstract (Originalsprache)
Background
Digital auscultatory blood pressure (BP) capture systems record Korotkoff sounds across a wide range of signal loudness levels. Whether signal loudness influences inter-rater agreement between trained reviewers has not been evaluated. If loudness were a meaningful predictor of disagreement, volume-based record screening would be warranted. Hypothesis: Loudness, measured using perceptually weighted integrated loudness (LUFS — Loudness Units Full Scale, per ITU-R BS.1770-4), does not predict inter-rater auscultatory BP agreement between trained observers.
Methods
Historical BP verification data from 341 subjects (177 female, 143 male; age 57±15 years) contributing 11,718 recordings were analyzed. Integrated loudness (LUFS) and loudness range (LR, in loudness units) were computed over two windows per recording: the full signal and during only the oscillometric deflation envelope. Correlations were computed between each volume metric and the absolute inter-reviewer difference in systolic and diastolic BP across 8,700 paired recordings reviewed independently by two trained observers. Logistic regression and receiver-operating-characteristic area under the curve (AUC) evaluated whether LUFS discriminated recordings where reviewers diverged by at least 4 mmHg (the ISO 81060-2 maximum allowable inter-rater difference). Bootstrap stability of signal loudness across ISO-sampled subsets was assessed over 200 replications at five sample sizes (25 to 100 subjects).
Results
Full-signal LUFS ranged widely across recordings (mean -29.5 dBFS; 5th to 95th percentile -56.8 to -13.1 dBFS, a 43 dBFS span), confirming that the study population spans a clinically broad loudness range. All volume-to-agreement correlations were negligible: the largest observed effect was an absolute Pearson r of 0.073 (LUFS versus absolute systolic difference). The AUC for LUFS predicting reviewer divergence at the ISO 4 mmHg threshold was approximately 0.48, indistinguishable from chance. Results were consistent whether computed over the full signal or restricted to the deflation envelope. Volume distributions across ISO-sampled subsets were stable, with the standard deviation of the sample mean LUFS not exceeding 2.0 dBFS across all sample sizes tested.
Conclusions
Audio signal loudness does not affect the accuracy of blood pressure determination, as demonstrated by inter-rater auscultatory BP disagreement across a wide dynamic range of 43 dBFS; a range that represents volume levels in a quiet room to the volume of sounds on a busy street. Near-chance discrimination at clinically motivated thresholds, confirmed by both correlation analysis and AUC, indicates that loudness-based record screening would not reduce inter-reviewer agreement. These findings support the use of post-hoc digital auscultation across naturally occurring signal levels in reference BP measurement studies.
Zitieren
APA
Sinha, T., & Alpert, B. S. (2026). Abstract FR426: Loudness Does Not Predict Inter-Rater Blood Pressure Agreement in Digitally Captured Auscultatory Recordings. Hypertension, 83(Suppl_1). https://doi.org/10.1161/hyp.83.suppl_1.fr426
Vancouver
Sinha T, Alpert BS. Abstract FR426: Loudness Does Not Predict Inter-Rater Blood Pressure Agreement in Digitally Captured Auscultatory Recordings. Hypertension. 2026;83(Suppl_1). doi:10.1161/hyp.83.suppl_1.fr426
BibTeX
@article{Sinha2026,
author = {Sinha, Tuhin and Alpert, Bruce S},
title = {Abstract FR426: Loudness Does Not Predict Inter-Rater Blood Pressure Agreement in Digitally Captured Auscultatory Recordings},
journal = {Hypertension},
year = {2026},
volume = {83},
number = {Suppl_1},
doi = {10.1161/hyp.83.suppl_1.fr426}
}
EUVIMED