Background
Dual-observer auscultatory blood pressure (BP) is the reference standard for sphygmomanometer verification under international standards. Inter-rater agreement across independent sites and study designs has not been systematically characterized using modern individual-level metrics. Hypothesis Dual-observer auscultatory BP demonstrates consistently high inter-rater agreement across independent studies when assessed with population-level and individual-level metrics.
Methods
Paired simultaneous auscultatory systolic (S) and diastolic (D) BP readings from three independent datasets were analyzed: an aggregated single-site dataset with 1,908 subjects across 52 historical verification studies (6,907 S and 6,903 D pairs); a prospective verification study at a site with 202 subjects (792 S and 791 D pairs); and a second study at that site with 185 subjects (730 S and 730 D pairs). Paired readings, from two simultaneous manual auscultations were analyzed in all of the datasets. Six agreement metrics were computed per BP component: bias plus standard deviation; 95% limits of agreement; concordance correlation coefficient (CCC); intraclass correlation coefficient (ICC); total deviation index (TDI); coverage probability at delta equals 5 mmHg (CP); and the Coefficient of Individual Agreement (CIA).
Results
Near-zero systematic bias was observed across all datasets (S: +0.07 to +0.44 mmHg; D: -0.27 to 0.00 mmHg). Concordance remained high across cohorts, with CCC at least 0.99 for all components (S range 0.9969 to 0.9984; D range 0.9908 to 0.9946). Individual-level agreement was strong, with TDI ranging from 2.68 to 3.76 mmHg for S and 2.94 to 4.25 mmHg for D. CP at delta equals 5 mmHg exceeded 0.978 for every dataset-component pair. CIA values were greater than 1 in all groups (from 1.38 to 1.88), consistent with cross-observer disagreement being no larger than within-observer replicate variability.
Conclusions
Dual-observer auscultatory BP demonstrates near-perfect inter-rater agreement across independent clinical sites and study designs spanning more than a decade. Replicate-aware individual agreement analysis shows observer-to-observer variability remains at or below within-observer replicate variability. These findings support dual-observer auscultatory measurement as a robust reference standard for BP device verification and establish practical baseline expectations for measurable accuracy in real-world verification studies.
Zitieren
APA
Sinha, T., Alpert, B. S., Grim, C. M., & Grim, C. E. (2026). Abstract FR425: Inter-Rater Auscultatory Blood Pressure Agreement Across Independent Clinical Studies. Hypertension, 83(Suppl_1). https://doi.org/10.1161/hyp.83.suppl_1.fr425
Vancouver
Sinha T, Alpert BS, Grim CM, Grim CE. Abstract FR425: Inter-Rater Auscultatory Blood Pressure Agreement Across Independent Clinical Studies. Hypertension. 2026;83(Suppl_1). doi:10.1161/hyp.83.suppl_1.fr425
BibTeX
@article{Sinha2026,
author = {Sinha, Tuhin and Alpert, Bruce S and Grim, Carlene M and Grim, Clarence E},
title = {Abstract FR425: Inter-Rater Auscultatory Blood Pressure Agreement Across Independent Clinical Studies},
journal = {Hypertension},
year = {2026},
volume = {83},
number = {Suppl_1},
doi = {10.1161/hyp.83.suppl_1.fr425}
}
EUVIMED