Multicenter Evaluation of the Feasibility of Clinical Implementation of SPECT Myocardial Blood Flow Measurement: Intersite Variability and Imaging Time
Circulation: Cardiovascular Imaging · 2023
DOI: 10.1161/circimaging.122.015009 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Circulation: Cardiovascular Imaging, 2023. Verfasst von R Glenn Wells und 13 weiteren Autor:innen. Beteiligte Einrichtung: Cardiology, University of Ottawa Heart Institute, Ottawa, Canada (R.G.W., S.G.C.-W., L.H., Y.K., T.D.R.). Hauptthemen laut MEDLINE: Coronary Artery Disease, Myocardial Perfusion Imaging. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 35 Literaturangaben.
Abstract (Originalsprache)
Background
Single-center studies have shown that single photon emission computed tomography myocardial blood flow (MBF) measurement is accurate compared with MBF measured with microspheres in a porcine model, positron emission tomography, and angiography. Clinical implementation requires consistency across multiple sites. The study goal is to determine the intersite processing repeatability of single photon emission computed tomography MBF and the additional camera time required.
Methods
Five sites (Canada, Italy, Japan, Germany, and Singapore) each acquired 25 to 35 MBF studies at rest and with pharmacological stress using technetium-99m-tetrofosmin on a pinhole-collimated cadmium-zinc-telluride–based cardiac single photon emission computed tomography camera with standardized list-mode imaging and processing protocols. Patients had intermediate to high pretest probability of coronary artery disease. MBF was measured locally and at a core laboratory using commercially available software. The time a room was occupied for an MBF study was compared with that for a standard rest/stress myocardial perfusion study.
Results
With motion correction, the overall correlation in MBF between core laboratory and local site was 0.93 (range, 0.87–0.97) at rest, 0.90 (range, 0.84–0.96) at stress, and 0.84 (range, 0.70–0.92) for myocardial flow reserve. The local-to-core difference in global MBF (bias -MBF ) was 5.4% (−3.8% to 14.8%; median [interquartile range]) at rest and 5.4% (−6.2% to 19.4%) at stress. Between the 5 sites, bias -MBF ranged from −1.6% to 11.0% at rest and from −1.9% to 16.3% at stress; the interquartile range in bias -MBF was between 9.3% (4.8%–14.0%) and 22.3% (−10.3% to 12.0%) at rest and between 17.0% (−11.3% to 5.6%) and 33.3% (−10.4% to 22.9%) at stress and was not significantly different between most sites. Both bias and interquartile range were like previously reported interobserver variability and less than the SD of the test-retest difference of 30%. The overall difference in myocardial flow reserve was 1.52% (−10.6% to 11.3%). There were no significant differences between with and without motion correction. The average additional acquisition time varied between sites from 44 to 79 minutes.
Conclusions
The average bias -MBF and bias -MFR values were small with standard deviations substantially less than the test-retest variability. This demonstrates that MBF can be measured consistently across multiple sites and further supports that this technique can be reliably implemented. REGISTRATION: URL: https://www.clinicaltrials.gov ; Unique identifier: NCT03427749.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cardiology, University of Ottawa Heart Institute, Ottawa, Canada (R.G.W., S.G.C.-W., L.H., Y.K., T.D.R.).
- Medizinische Hochschule Hannover, Hannover, Germany (F.M.B., J.D.).
- Nuclear Medicine, Università & Spedali Civili, Brescia, Italy (L.C., E.C.).
- Department of Radiology, Ehime University Graduate School of Medicine, Ehime, Japan (N.K., M.M.).
- Cardiology, National Heart Center Singapore, Singapore (Y.J.F.K., X.F.T.).
- GE Healthcare, Haifa, Israel (H.R.).
ORCID-Profile: R Glenn Wells, Frank M Bengel, Luca Camoni, Elisabetta Cerudelli, Johanna Diekmann, Yoshito Kadoya, Masao Miyagawa, Terrence D Ruddy
Zitieren
APA
Wells, R. G., Bengel, F. M., Camoni, L., Cerudelli, E., Cuddy-Walsh, S. G., Diekmann, J., Han, L., Kadoya, Y., Kawaguchi, N., Keng, Y. J. F., Miyagawa, M., Ratner, H., Teng, X. F., & Ruddy, T. D. (2023). Multicenter Evaluation of the Feasibility of Clinical Implementation of SPECT Myocardial Blood Flow Measurement: Intersite Variability and Imaging Time. Circulation: Cardiovascular Imaging. https://doi.org/10.1161/circimaging.122.015009
Vancouver
Wells RG, Bengel FM, Camoni L, Cerudelli E, Cuddy-Walsh SG, Diekmann J, et al. Multicenter Evaluation of the Feasibility of Clinical Implementation of SPECT Myocardial Blood Flow Measurement: Intersite Variability and Imaging Time. Circulation: Cardiovascular Imaging. 2023. doi:10.1161/circimaging.122.015009
BibTeX
@article{Wells2023,
author = {Wells, R Glenn and Bengel, Frank M and Camoni, Luca and Cerudelli, Elisabetta and Cuddy-Walsh, Sarah G and Diekmann, Johanna and Han, Lewis and Kadoya, Yoshito and Kawaguchi, Naoto and Keng, Yung Jih Felix and Miyagawa, Masao and Ratner, Hagit and Teng, Xue Fen and Ruddy, Terrence D},
title = {Multicenter Evaluation of the Feasibility of Clinical Implementation of SPECT Myocardial Blood Flow Measurement: Intersite Variability and Imaging Time},
journal = {Circulation: Cardiovascular Imaging},
year = {2023},
doi = {10.1161/circimaging.122.015009}
}
EUVIMED