EUVIMEDEUVIMED entdecken
Vergleichsstudie

Comparison of Three-Dimensional Accuracy of Conventional Cast Metal to Milled Zirconia Implant Full-Arch Fixed Dental Prosthesis Frameworks

Eunice Siew Pei Lua, Keson Beng Choon Tan, Frank Kong Fei Lee, Sophia Hui Xin Yee, Keng Mun Wong, Ming Yi Tan

The International Journal of Oral & Maxillofacial Implants · 2025

DOI: 10.11607/jomi.11262 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift The International Journal of Oral & Maxillofacial Implants, 2025. Verfasst von Eunice Siew Pei Lua und 5 weiteren Autor:innen. Hauptthemen laut MEDLINE: Zirconium, Dental Prosthesis, Implant-Supported, Imaging, Three-Dimensional, Denture Design.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: To compare the 3D accuracy of conventional and digital workflows in the fabrication of implant-supported full-arch fixed dental prosthesis (FAFDP) frameworks for an edentulous mandible. MATERIALS AND METHODS: A heat-polymerized polymethyl methacrylate (PMMA) master model simulated an edentulous mandible to be restored with a fixed prosthesis supported by five implants (implants A to E). A conventional open-tray, splinted-coping polyether impression was poured with type 4 dental stone to produce a stone model, which was scanned with a dental laboratory scanner to produce a virtual model. Zirconia (Zr) full-contour (ZFC) (n = 5) and Zr cutback (ZCB) (n = 5) frameworks were fabricated with CAD/CAM. The stone model was used as a working model to fabricate conventional noble metal frameworks (CNB) (n = 5). Each test framework was attached to five implants with application of 35-Ncm torque. Test models were fabricated with type 4 dental stone and were allowed to set completely before removing the test frameworks. Centroid positions and central axes of the implants in master and test models were measured with a coordinate measuring machine and compared. The local coordinate system comprised implant A (right posterior implant) as the origin; implants A, C (central incisor implant), and E (left posterior implant) as the xy-plane; and implants A and E as the x-axis. Three linear distortion parameters (dx, dy, and dz), one global linear distortion parameter (dR), two angular distortion parameters (dθx and dθy), and 3D distance distortion values (ΔR and %ΔR) were assessed. RESULTS: The magnitude of mean dx, dy, dz, and dR values ranged from 39.9 ± 26.4 µm (CNB-B) to 263.3 ± 94.6 µm (CNB-E) for dx; from 56.9 ± 5.4 µm (ZFC-B) to 124.5 ± 37.6 µm (ZCB-C) for dy; from -8.3 ± 27.6 µm (ZFC-D) to -24.6 ± 17.3 µm (ZCB-D) for dz; and from 80.6 ± 29.4 µm (CNB-B) to 263.3 ± 94.6 µm (CNB-E) for dR. The magnitude of mean dθx and dθy ranged from 0.025 ± 0.486 degrees (ZFC-C) to 1.490 ± 0.383 degrees (CNB-C) and from -0.050 ± 0.171 degrees (ZCB-B) to 1.263 ± 0.501 degrees (ZFC-C), respectively. One-way analysis of variance (ANOVA) found differences among groups for dx, dy, dz, dθx, and dθy at some implants, but there was no clear pattern on the poorest group. For global linear distortion (dR), CNB-D fared the worst. However, CNB exhibited the worst accuracy in ΔR and %ΔR for all reference distances, except implant A to implant E. CONCLUSIONS: There were no differences between ZFC and ZCB for all parameters. CNB fared poorest for ΔR and %ΔR for three of the four reference distances (A-B, A-C, and A-D), as well as for dR for one of five implant positions (implant D).

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ZirconiumDental Prosthesis, Implant-SupportedImaging, Three-DimensionalDenture DesignDental Prosthesis DesignHumansComputer-Aided DesignModels, DentalJaw, EdentulousMandibleDental Impression TechniqueDental Implants

Stichwörter der Autor:innen

3D accuracyCAD/CAMcoordinate measuring machineimplant prosthodontics

Substanzen

Zirconiumzirconium oxideDental Implants

Zitieren

APA

Lua, E. S. P., Tan, K. B. C., Lee, F. K. F., Yee, S. H. X., Wong, K. M., & Tan, M. Y. (2025). Comparison of Three-Dimensional Accuracy of Conventional Cast Metal to Milled Zirconia Implant Full-Arch Fixed Dental Prosthesis Frameworks. The International Journal of Oral & Maxillofacial Implants. https://doi.org/10.11607/jomi.11262

Vancouver

Lua ESP, Tan KBC, Lee FKF, Yee SHX, Wong KM, Tan MY. Comparison of Three-Dimensional Accuracy of Conventional Cast Metal to Milled Zirconia Implant Full-Arch Fixed Dental Prosthesis Frameworks. The International Journal of Oral & Maxillofacial Implants. 2025. doi:10.11607/jomi.11262

BibTeX
@article{Lua2025,
  author = {Lua, Eunice Siew Pei and Tan, Keson Beng Choon and Lee, Frank Kong Fei and Yee, Sophia Hui Xin and Wong, Keng Mun and Tan, Ming Yi},
  title = {Comparison of Three-Dimensional Accuracy of Conventional Cast Metal to Milled Zirconia Implant Full-Arch Fixed Dental Prosthesis Frameworks},
  journal = {The International Journal of Oral & Maxillofacial Implants},
  year = {2025},
  doi = {10.11607/jomi.11262}
}
Kostenlos starten →