Effect of Lowering Dialysate Flow Rate on Dialysis Dose and Laboratory Values
Blood Purification · 2025
DOI: 10.1159/000545898 ↗Kurz gefasst
Fachartikel in der Zeitschrift Blood Purification, 2025. Verfasst von Maryn Gardner und 7 weiteren Autor:innen. Beteiligte Einrichtung: Division of Nephrology and Hypertension, Department of Internal Medicine, Vanderbilt University Medical Center, Nashville, Tennessee, USA. Hauptthemen laut MEDLINE: Renal Dialysis, Dialysis Solutions. Die Arbeit nennt 10 Literaturangaben.
Abstract (Originalsprache)
Introduction
The rationale behind using higher dialysate flow rates stems from the idea of maximizing a dialysis membrane’s extraction ratio. Subsequent developments in dialysis membrane design, including the hollow fiber undulations, spacer yarns, and altering the fiber packing density, helped reduce blood and dialysate flow mismatch in dialysis membranes.
Methods
This is a single-center retrospective analysis of in-center dialysis patients. Adult (age ≥18 years) in-center hemodialysis patients who were receiving dialysis at the centers 3 months prior to and 3 months following the change in dialysate flow rates were included in the study (December 2022 to May 2023). Dialysate flow rates were reduced from 800 to 600 mL/min – using the same access and dialysis membrane.
Results
Sixty-eight patients were enrolled in the study. The change of dialysate flow rate did not result in a statistically significant change in dialysis dose (Kt/Vurea). All regression coefficients were below 0.05 (i.e., a 0.05 change in dialysis dose for flow rate 600 vs. 800 mL/min). Significant associations were found between dialysis dose and both the blood flow rate and dialysate sodium, but these were also clinically negligible differences. Analysis of patients’ laboratory testing showed no statistically significant difference in serum electrolyte values (sodium, potassium, bicarbonate, blood urea nitrogen, and creatinine).
Conclusion
In patients treated using the same dialysate filter and dialysis access, changing the dialysate flow rate from 800 to 600 mL/min does not result in clinically significant changes in either patients’ measured serum electrolytes or dialysis dose.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Division of Nephrology and Hypertension, Department of Internal Medicine, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
- Division of Nephrology and Hypertension, Department of Medicine, University of North Carolina at Chapel Hill, Chapel Hill, North Carolina, USA.
- Department of Surgery, George Washington Transplant Institute, George Washington University Hospital, Washington, DC, USA.
- Department of Biostatistics, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
- Center for Health Services Research, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
- Division of Renal Diseases and Hypertension, George Washington University, Washington, DC, USA.
Zitieren
APA
Gardner, M., Wyatt, N. E., Sanghani, N., Attia, A., Guide, A., Greevy, R. A., Salani, M., & El Shamy, O. (2025). Effect of Lowering Dialysate Flow Rate on Dialysis Dose and Laboratory Values. Blood Purification. https://doi.org/10.1159/000545898
Vancouver
Gardner M, Wyatt NE, Sanghani N, Attia A, Guide A, Greevy RA, et al. Effect of Lowering Dialysate Flow Rate on Dialysis Dose and Laboratory Values. Blood Purification. 2025. doi:10.1159/000545898
BibTeX
@article{Gardner2025,
author = {Gardner, Maryn and Wyatt, Nicole E and Sanghani, Neil and Attia, Ahmed and Guide, Andrew and Greevy, Robert A and Salani, Megha and El Shamy, Osama},
title = {Effect of Lowering Dialysate Flow Rate on Dialysis Dose and Laboratory Values},
journal = {Blood Purification},
year = {2025},
doi = {10.1159/000545898}
}
EUVIMED