Effective Radiation Dose from Cone-Beam Computed Tomography Guidance during Bronchoscopic Tumour Ablation

Jack Mitchell Watson, Paul F Einsiedel, Phillip Antippa, Kanishka Rangamuwa, Louis Irving, Daniel P Steinfort

Respiration · 2024

DOI: 10.1159/000539862 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Respiration, 2024. Verfasst von Jack Mitchell Watson und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Respiratory and Sleep Medicine, Royal Melbourne Hospital, Melbourne, Victoria, Australia. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Bronchoscopy, Lung Neoplasms, Carcinoma, Non-Small-Cell Lung. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 20 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Endobronchial radiofrequency ablation (RFA) is a novel minimally invasive approach to management of peripheral non-small-cell lung cancer (NSCLC) in medically inoperable patients. Minimally invasive ablative techniques are generally delivered with cone-beam computed tomography (CBCT) guidance. CBCT requires a significant number of two dimensional imaging projections to be acquired which is then reconstructed as a three-dimensional cone-beam image. The objective of this study was to determine the radiation dosimetry consequent to use of CBCT guidance for bronchoscopic RFA.

Methods

Post hoc analysis of data following bronchoscopic RFA of stage I biopsy-confirmed NSCLC performed with CBCT. Effective dose estimates for these patients were calculated using PCXMC2.0 software.

Results

Ten patients underwent bronchoscopic RFA, with a median 3 (range 2–4) CBCT spins per procedure. Mean dose area product (DAP) per procedure was 7,778 μGy.m2 (±4,743) with an effective dose of 11.6 mSv (±7.4). The DAP per spin for these 10 patients varied from 83.8 to 8,625.6 μGy.m2 (effective dose range 0.15–13.81 mSv).

Conclusion

This is the first study to report radiation dosimetry consequent to CT guidance for bronchoscopic RFA procedures. Effective doses appear comparable to other CT fluoroscopic procedures.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cone-Beam Computed TomographyBronchoscopyLung NeoplasmsCarcinoma, Non-Small-Cell LungRadiation DosageHumansMaleFemaleAgedMiddle AgedRadiofrequency AblationAged, 80 and overRetrospective Studies

Stichwörter der Autor:innen

BronchoscopyInterventional pulmonologyLung cancerRadiation

Autor:innen und Institutionen

  • Department of Respiratory and Sleep Medicine, Royal Melbourne Hospital, Melbourne, Victoria, Australia.
  • Department of Medical Imaging, Royal Melbourne Hospital, Melbourne, Victoria, Australia.
  • Cardiothoracic Surgery Unit, Royal Melbourne Hospital, Melbourne, Victoria, Australia.
  • Department of Surgery, University of Melbourne, Melbourne, Victoria, Australia.
  • Department of Medicine, University of Melbourne, Melbourne, Victoria, Australia.

Zitieren

APA

Watson, J. M., Einsiedel, P. F., Antippa, P., Rangamuwa, K., Irving, L., & Steinfort, D. P. (2024). Effective Radiation Dose from Cone-Beam Computed Tomography Guidance during Bronchoscopic Tumour Ablation. Respiration. https://doi.org/10.1159/000539862

Vancouver

Watson JM, Einsiedel PF, Antippa P, Rangamuwa K, Irving L, Steinfort DP. Effective Radiation Dose from Cone-Beam Computed Tomography Guidance during Bronchoscopic Tumour Ablation. Respiration. 2024. doi:10.1159/000539862

BibTeX
@article{Watson2024,
  author = {Watson, Jack Mitchell and Einsiedel, Paul F and Antippa, Phillip and Rangamuwa, Kanishka and Irving, Louis and Steinfort, Daniel P},
  title = {Effective Radiation Dose from Cone-Beam Computed Tomography Guidance during Bronchoscopic Tumour Ablation},
  journal = {Respiration},
  year = {2024},
  doi = {10.1159/000539862}
}
Kostenlos starten →