Efficacy of a Novel Medical Adhesive for Sealing Lung Parenchyma: An in vitro Study in Rabbit Lungs
European Surgical Research · 2021 · 62 (4) · 255-261
DOI: 10.1159/000517173 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Surgical Research, online erschienen am 21.07.2021. Verfasst von Sebastian Kalverkamp und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Thoracic and Cardiovascular Surgery, University Hospital RWTH Aachen, Aachen, Germany. Hauptthemen laut MEDLINE: Adhesives, Lung, Parenchymal Tissue. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 18 Literaturangaben.
Abstract (Originalsprache)
Introduction
During thoracic resection procedures, complete hemostasis and aerostasis are priorities. A persistent alveolar air leak is associated with increased morbidity and mortality rates. This study aimed to evaluate whether the novel medical adhesive VIVO (Adhesys Medical GmbH Aachen, Germany) is a reliable alternative sealing technique to routine surgical procedures.
Methods
We conducted an in vitro animal study by analyzing 21 lungs of New Zealand (n = 19) and Chinchilla Bastard (n = 2) rabbits (age, 11–18 weeks; weight, 2,400–3,600 g). Three groups, each comprising 7 animals, were evaluated. VIVO (VIVO-group) was compared with standard surgical lung parenchymal lesion closure with a polypropylene suture (Suture-group) and TachoSil® (TachoSil-group). We adopted a stable, pressure-controlled ventilation protocol. After explantation, a surgical incision 0.5-cm deep and 1.5-cm wide was made in the lungs using a customized template. Air leak was measured quantitatively (mL/min) using a respirator and visualized qualitatively by 2 observers who made independent judgments. Next, the leak was closed using VIVO, suture, or TachoSil® as specified by the manufacturer. Subsequently, positive end-expiratory pressure (PEEP) and inspiratory pressure were gradually increased until a maximum of 15 and 30 mbar were attained, respectively.
Results
At PEEPs of 8, 10, and 15 mbar, VIVO achieved complete sealing of the profound parenchymal defect in all (n = 7) lungs. After closure of the incision, we observed an air leak variation of 127 ± 114 mL/min (Suture-group), 31 ± 49 mL/min (VIVO-group), and 114 ± 134 mL/min (TachoSil-group). VIVO showed a significantly lower air leak than surgical sutures (p = 0.031) and TachoSil® (p = 0.046).
Conclusion
VIVO offers sufficient closure of the lung parenchymal lesions. The novel adhesive enabled significantly better sealing with lower persistent air leakage than TachoSil® or surgical sutures. Further investigation using in vivo models is strongly encouraged to confirm our findings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Thoracic and Cardiovascular Surgery, University Hospital RWTH Aachen, Aachen, Germany.
- Adhesys Medical GmbH, Aachen, Germany.
- Institute for Laboratory Animal Science & Experimental Surgery, RWTH Aachen, University Hospital RWTH Aachen, Aachen, Germany.
Zitieren
APA
Kalverkamp, S., Mantas, A., Spillner, J., Hima, F., Kanzler, S. S., Stopinski, T., Tolba, R. H., & Zayat, R. (2021). Efficacy of a Novel Medical Adhesive for Sealing Lung Parenchyma: An in vitro Study in Rabbit Lungs. European Surgical Research, 62(4), 255-261. https://doi.org/10.1159/000517173
Vancouver
Kalverkamp S, Mantas A, Spillner J, Hima F, Kanzler SS, Stopinski T, et al. Efficacy of a Novel Medical Adhesive for Sealing Lung Parenchyma: An in vitro Study in Rabbit Lungs. European Surgical Research. 2021;62(4):255-261. doi:10.1159/000517173
BibTeX
@article{Kalverkamp2021,
author = {Kalverkamp, Sebastian and Mantas, Anna and Spillner, Jan and Hima, Flutura and Kanzler, Stephanie Sarah and Stopinski, Thaddäus and Tolba, René H and Zayat, Rashad},
title = {Efficacy of a Novel Medical Adhesive for Sealing Lung Parenchyma: An in vitro Study in Rabbit Lungs},
journal = {European Surgical Research},
year = {2021},
volume = {62},
number = {4},
pages = {255--261},
doi = {10.1159/000517173}
}
EUVIMED