Intraoperative Correction of Cochlear Implant Electrode Translocation
Audiology and Neurotology · 2022 · 27 (2) · 104-108
DOI: 10.1159/000515684 ↗Kurz gefasst
Fachartikel in der Zeitschrift Audiology and Neurotology, online erschienen am 29.04.2021. Verfasst von William G Morrel und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otolaryngology-Head and Neck Surgery, Vanderbilt University Medical Center, Nashville, Tennessee, USA. Hauptthemen laut MEDLINE: Cochlear Implantation, Cochlear Implants. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 12 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIDCD NIH HHS.
Abstract (Originalsprache)
Introduction
Translocation of precurved cochlear implant (CI) electrodes reduces hearing outcomes, but it is not known whether it is possible to correct scalar translocation such that all electrodes reside fully in the scala tympani (ST).
Methods
Six cadaveric temporal bones were scanned with CT and segmented to delineate intracochlear anatomy. Mastoidectomy with facial recess was performed. Precurved CI electrodes (CI532; Cochlear Limited) were implanted until scalar translocation was confirmed with postoperative CT. Then, electrodes were removed and replaced. CT scan was repeated to assess for translocation correction. Scalar position of electrode contacts, angular insertion depth (AID) of the electrode array, and M– (average distance between each electrode contact and the modiolus) were measured. An in vivo case is reported in which intraoperative translocation detection led to removal and replacement of the electrode.
Results
Five of 6 cadaveric translocations (83%) were corrected with 1 attempt, resulting in full ST insertions. AID averaged 285 ± 77° for translocated electrodes compared to 344 ± 28° for nontranslocated electrodes (p = 0.109). M– averaged 0.75 ± 0.18 mm for translocated electrodes and 0.45 ± 0.11 mm for nontranslocated electrodes (p = 0.016). Reduction in M– with translocation correction averaged 38%. In the in vivo case, translocation was successfully corrected in a single attempt.
Conclusion
Scalar translocation of precurved CI electrodes can be corrected by removal and reinsertion. This significantly improves the perimodiolar positioning of these electrodes. There was a high rate of success (83%) in this cadaveric model as well as a successful in vivo attempt.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Otolaryngology-Head and Neck Surgery, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
- Department of Electrical Engineering and Computer Science, Vanderbilt University, Nashville, Tennessee, USA.
- Department of Biomedical Engineering, Vanderbilt University, Nashville, Tennessee, USA.
Förderung
- NIDCD NIH HHS
Zitieren
APA
Morrel, W. G., Manzoor, N. F., Dawant, B. M., Noble, J. H., & Labadie, R. F. (2022). Intraoperative Correction of Cochlear Implant Electrode Translocation. Audiology and Neurotology, 27(2), 104-108. https://doi.org/10.1159/000515684
Vancouver
Morrel WG, Manzoor NF, Dawant BM, Noble JH, Labadie RF. Intraoperative Correction of Cochlear Implant Electrode Translocation. Audiology and Neurotology. 2022;27(2):104-108. doi:10.1159/000515684
BibTeX
@article{Morrel2022,
author = {Morrel, William G and Manzoor, Nauman F and Dawant, Benoit M and Noble, Jack H and Labadie, Robert F},
title = {Intraoperative Correction of Cochlear Implant Electrode Translocation},
journal = {Audiology and Neurotology},
year = {2022},
volume = {27},
number = {2},
pages = {104--108},
doi = {10.1159/000515684}
}
EUVIMED