A Continuous and Non-Invasive Arterial Pressure Monitoring System in Dialysis Patients
Nephron Clinical Practice · 01.11.2007 · 107 (4) · c170-c176
DOI: 10.1159/000110677 ↗Kurz gefasst
Fachartikel in der Zeitschrift Nephron Clinical Practice, online erschienen am 01.11.2007. Verfasst von Emanuele Mambelli und 2 weiteren Autor:innen. Beteiligte Einrichtung: Nephrology, Dialysis and Hypertension Unit, Policlinico S. Orsola-Malpighi, Bologna, Italy. Hauptthemen laut MEDLINE: Blood Pressure Monitors. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 11 Literaturangaben.
Abstract (Originalsprache)
Background
As symptomatic intradialytic hypotension in the hemodialysis (HD) patient is often a sudden event whose onset cannot be predicted by means of extemporary measures, continuous blood pressure (BP) measurement would be far more useful. We tested a new continuous noninvasive BP monitoring system, Harmonized Alert Sensing Technology (HASTE), which, by means of the analysis of the finger pulse wave, obtained from an O2 sensor, estimates a beat-to-beat systolic pressure value (Esys) and supplies a continuous read-out. The study aim sought to verify the reliability of this non-invasive instrument in continuously providing systolic pressure values during HD.
Methods
We studied 18 patients during HD treatment, initially comparing the Esys with the invasive blood pressure measurement (SYS). Subsequently, the Esys derived from the O2 sensor in the arm with the shunt (S) and the arm without (N), respectively, were both compared with the cuff measurement.
Results
The mean difference between SYS and Esys was 0.7 ± 13.3 mm Hg (p < 0.01; r = 0.80). There was a mean difference of 0.2 ± 21.9 mm Hg (p = NS; r = 0.67) between Esys(N) and Esys(S). The correlation was not statistically significant even between Esys(N) or Esys(S) versus the cuff measurement, respectively. Patient movement compromised the accuracy of the estimations made using the O2 sensor in the non-fistula arm.
Conclusions
A good correlation between the data estimated by HASTE compared with invasive BP suggests that the instrument may prove useful for continuously monitoring the blood pressure trends during the dynamic hemodialysis situation. However, its sensibility needs to be improved in order to be used indifferently in both arms with a view to achieving real intradialytic hypotension prevention.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Nephrology, Dialysis and Hypertension Unit, Policlinico S. Orsola-Malpighi, Bologna, Italy.
Zitieren
APA
Mambelli, E., Mancini, E., & Santoro, A. (2007). A Continuous and Non-Invasive Arterial Pressure Monitoring System in Dialysis Patients. Nephron Clinical Practice, 107(4), c170-c176. https://doi.org/10.1159/000110677
Vancouver
Mambelli E, Mancini E, Santoro A. A Continuous and Non-Invasive Arterial Pressure Monitoring System in Dialysis Patients. Nephron Clinical Practice. 2007;107(4):c170-c176. doi:10.1159/000110677
BibTeX
@article{Mambelli2007,
author = {Mambelli, Emanuele and Mancini, Elena and Santoro, Antonio},
title = {A Continuous and Non-Invasive Arterial Pressure Monitoring System in Dialysis Patients},
journal = {Nephron Clinical Practice},
year = {2007},
volume = {107},
number = {4},
pages = {c170--c176},
doi = {10.1159/000110677}
}
EUVIMED