EUVIMEDEUVIMED entdecken
Validierungsstudie

Validation and subgroup analysis of the accuracy of the photoplethysmography-based Microlife cuffless upper-arm wearable blood pressure monitor

Ching-Fu Wang iD, Han-Lin Wang, Pei-Hsin Kuo, Shih-Zhang Li, Bo-Wei Chen, Ssu-Ju Li, You-Yin Chen iD, Sheng-Huang Lin iD

American Journal of Hypertension · 2025

DOI: 10.1093/ajh/hpaf238 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift American Journal of Hypertension, 2025. Verfasst von Ching-Fu Wang und 7 weiteren Autor:innen. Beteiligte Einrichtung: School of Biomedical Engineering, College of Biomedical Engineering, Taipei Medical University, Taipei City, Taiwan. Hauptthemen laut MEDLINE: Photoplethysmography, Blood Pressure, Blood Pressure Monitors, Blood Pressure Determination. Gefördert von: Tzu Chi Research and Department, Ministry of Science and Technology.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND This study (1) validates the accuracy of the photoplethysmography-based Microlife cuffless upper-arm wearable blood pressure (BP) monitor according to the AAMI/ESH/ISO 81060-2:2018/Amd 1:2020 standard; (2) investigates the device’s performance across age groups and skin tone; and (3) evaluates the advantages of cuffless BP monitoring in terms of estimation variability, required time, and stability. METHODS A total of 120 participants aged ≥20 years were recruited from the general population, ensuring the representation of diverse age and skin tone groups. The device was calibrated against a clinically validated mercury sphygmomanometer. Sequential same-arm BP estimation were performed. Accuracy was analyzed according to the ISO criteria, and further subgroup analysis was conducted to compare results between participants aged ≥65 years and <65 years, as well as between participants with lighter and darker skin tones (classified according to the Fitzpatrick system). Additionally, estimation stability and variability were evaluated using three averaged BP estimations. RESULTS The results demonstrated that the Microlife device met the AAMI/ESH/ISO accuracy criteria. Subgroup analysis revealed consistent accuracy across age and skin tone groups, with slight differences warranting further exploration. Notably, the cuffless design enabled faster estimations with reduced variation between the three averaged readings, showcasing its potential for home BP monitoring and frequent self-assessments. CONCLUSIONS These findings support the clinical potential of photoplethysmography-based cuffless BP monitoring in diverse adult populations, particularly for facilitating rapid and stable BP estimations in elderly individuals and users with varying skin tones. Further large-scale studies are warranted to corroborate and build upon these observations.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PhotoplethysmographyBlood PressureBlood Pressure MonitorsBlood Pressure DeterminationWearable Electronic DevicesHumansMaleFemaleMiddle AgedAgedAdultReproducibility of ResultsPredictive Value of TestsYoung AdultEquipment DesignAge Factors

Stichwörter der Autor:innen

cuffless blood pressure monitorestimation variabilityphotoplethysmography (PPG)skin tone and age subgroup analysisvalidation study

Autor:innen und Institutionen

  • School of Biomedical Engineering, College of Biomedical Engineering, Taipei Medical University, Taipei City, Taiwan.
  • Department of Biomedical Engineering, National Yang Ming Chiao Tung University, Taipei, Taiwan.
  • Department of Neurology, Hualien Tzu Chi Hospital, Buddhist Tzu chi Medical Foundation, Hualien, Taiwan.
  • Department of Neurology, School of Medicine, Tzu Chi University, Hualien, Taiwan.
  • Ph.D. Program in Medical Neuroscience, College of Medical Science and Technology, Taipei Medical University, New Taipei City, Taiwan.
  • Medical Device Innovation and Translation Center, National Yang Ming Chiao Tung University, Taipei, Taiwan.

ORCID-Profile: Ching-Fu Wang, You-Yin Chen, Sheng-Huang Lin

Förderung

  • Tzu Chi Research and Department
  • Ministry of Science and Technology

Zitieren

APA

Wang, C. F., Wang, H. L., Kuo, P. H., Li, S. Z., Chen, B. W., Li, S. J., Chen, Y. Y., & Lin, S. H. (2025). Validation and subgroup analysis of the accuracy of the photoplethysmography-based Microlife cuffless upper-arm wearable blood pressure monitor. American Journal of Hypertension. https://doi.org/10.1093/ajh/hpaf238

Vancouver

Wang CF, Wang HL, Kuo PH, Li SZ, Chen BW, Li SJ, et al. Validation and subgroup analysis of the accuracy of the photoplethysmography-based Microlife cuffless upper-arm wearable blood pressure monitor. American Journal of Hypertension. 2025. doi:10.1093/ajh/hpaf238

BibTeX
@article{Wang2025,
  author = {Wang, Ching-Fu and Wang, Han-Lin and Kuo, Pei-Hsin and Li, Shih-Zhang and Chen, Bo-Wei and Li, Ssu-Ju and Chen, You-Yin and Lin, Sheng-Huang},
  title = {Validation and subgroup analysis of the accuracy of the photoplethysmography-based Microlife cuffless upper-arm wearable blood pressure monitor},
  journal = {American Journal of Hypertension},
  year = {2025},
  doi = {10.1093/ajh/hpaf238}
}
Kostenlos starten →