Effects of Endourologic Devices and Connectors on Irrigation Flow Through the Flexible Ureteroscope Working Channel

Oleksandr Chepeliuk iD, Rinat Lasmanovich iD, Isaias Caballero Lopez, Dina Orkin, Miki Haifler iD, Dorit E Zilberman iD, Zohar A Dotan iD, Asaf Shvero iD, Nir Kleinmann

Advances in Urology · 2026-01 · 2026 (1)

DOI: 10.1155/aiu/7192796 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Advances in Urology, online erschienen am 30.07.2026. Verfasst von Oleksandr Chepeliuk und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Urology, Sheba Medical Center, Tel HaShomer, Ramat Gan, Israel, sheba.co.il. Die Arbeit nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To evaluate the effect of different connector types and commonly used endourologic devices on irrigation fluid flow through the working channel of a flexible ureteroscope.

Methods

Irrigation flow was measured under a 200‐cm H 2 O pressure head using three connector types (Tuohy‐Borst, UroLok, and Check‐Flo) and a standard Y‐set connection. Instruments tested included Piranha biopsy forceps, Segura hemisphere basket, Glidewire guidewire, and two laser fibers (200 μm and 365 μm). Each condition was repeated six times. Descriptive statistics were calculated, and one‐way ANOVA was used to assess differences between groups.

Results

With an empty channel, the standard Y‐set connection achieved the highest mean flow (115.8 ± 4.3 mL/min), followed by Tuohy‐Borst (113.0 ± 1.6), UroLok (100.2 ± 1.3), and Check‐Flo (80.4 ± 0.5) ( p < 0.001). Inserting instruments markedly reduced flow. The Piranha forceps and Glidewire guidewire produced the largest reduction (∼96–97%). A larger instrument diameter or the absence of a smooth coating significantly reduced flow across all connectors. With small‐caliber instruments, connector design played a greater role: relative flow reduction was greatest with Tuohy‐Borst (50.6%) and lowest with Check‐Flo (38.0%). In absolute terms, UroLok yielded the highest median flow with laser fibers (38.7 mL/min), followed by Tuohy‐Borst (37.5 mL/min) and Check‐Flo (34.7 mL/min).

Conclusion

Both connector choice and instrument size significantly affect irrigation flow in flexible ureteroscopy. Large‐diameter instruments almost abolish flow regardless of connector, whereas with smaller instruments, connector design influences performance. These findings emphasize the importance of selecting optimal connector–instrument combinations to maintain intraoperative visibility.

Schlagwörter

Stichwörter der Autor:innen

connector systemsflexible ureteroscopyirrigation flowlaser fibersureteral access sheath

Autor:innen und Institutionen

  • Department of Urology, Sheba Medical Center, Tel HaShomer, Ramat Gan, Israel, sheba.co.il.
  • Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel, tau.ac.il.
  • Operative Room Division, Anesthesia and Pain, Sheba Medical Center, Tel HaShomer, Ramat Gan, Israel, sheba.co.il.
  • Meir Medical Center, Kfar Saba, Israel, mmc.org.il.

ORCID-Profile: Oleksandr Chepeliuk, Rinat Lasmanovich, Miki Haifler, Dorit E Zilberman, Zohar A Dotan, Asaf Shvero

Zitieren

APA

Chepeliuk, O., Lasmanovich, R., Lopez, I. C., Orkin, D., Haifler, M., Zilberman, D. E., Dotan, Z. A., Shvero, A., & Kleinmann, N. (2026). Effects of Endourologic Devices and Connectors on Irrigation Flow Through the Flexible Ureteroscope Working Channel. Advances in Urology, 2026(1). https://doi.org/10.1155/aiu/7192796

Vancouver

Chepeliuk O, Lasmanovich R, Lopez IC, Orkin D, Haifler M, Zilberman DE, et al. Effects of Endourologic Devices and Connectors on Irrigation Flow Through the Flexible Ureteroscope Working Channel. Advances in Urology. 2026;2026(1). doi:10.1155/aiu/7192796

BibTeX
@article{Chepeliuk2026,
  author = {Chepeliuk, Oleksandr and Lasmanovich, Rinat and Lopez, Isaias Caballero and Orkin, Dina and Haifler, Miki and Zilberman, Dorit E and Dotan, Zohar A and Shvero, Asaf and Kleinmann, Nir},
  title = {Effects of Endourologic Devices and Connectors on Irrigation Flow Through the Flexible Ureteroscope Working Channel},
  journal = {Advances in Urology},
  year = {2026},
  volume = {2026},
  number = {1},
  doi = {10.1155/aiu/7192796}
}
Kostenlos starten →