Pilot Study on Postoperative Toric Intraocular Lens Alignment
Journal of Ophthalmology · 2024-01 · 2024 (1)
DOI: 10.1155/2024/1053914 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Ophthalmology, online erschienen am 14.11.2024. Verfasst von Baodi Yang und 6 weiteren Autor:innen. Beteiligte Einrichtung: The First School of Clinical Medicine, Southern Medical University, Guangzhou, China. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Natural Science Foundation of Guangdong Province, National Natural Science Foundation of China.
Abstract (Originalsprache)
Purpose
Assess the comparative accuracy and reliability of postoperative toric intraocular lens (TIOL) alignment measurement methods: Casia2 and Adobe Photoshop with digital slit lamp images (PS method).
Methods
In a study of 41 subjects with 58 eyes postimplantation of TIOL, we independently measured TIOL alignment with Casia2 and PS methods. The intraclass correlation coefficient (ICC 1,1 ) was employed to assess the repeatability of the Casia2 method. While ICC 2,1 for absolute agreement and Bland–Altman analysis were utilized to determine the interdevice agreement between the two methods, the regression analysis was conducted to identify any proportional bias.
Results
Casia2 demonstrated excellent intradevice repeatability with an ICC 1,1 of 0.998. The absolute agreement between Casia2 and PS was very high with an ICC 2,1 of 0.999. The average discrepancy between the two measurement methods was −0.828°, with a 95% confidence interval (CI) ranging from −1.623° to −0.032°. The 95% limits of agreement (LoA) were between −6.761° and 5.105°, indicating a strong concordance in TIOL alignment measurements. Casia2 was capable of accurately measuring the TIOL axis alignment under conditions of pupil diameters (PDs) of 4 mm or greater.
Conclusion
Casia2 and PS demonstrated significant concordance in measuring postoperative TIOL alignment, with Casia2 offering a more straightforward and accessible alternative, particularly beneficial for patients with suboptimal pupil dilation.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- The First School of Clinical Medicine, Southern Medical University, Guangzhou, China.
- Department of Ophthalmology, The Second Affiliated Hospital, School of Medicine, Shenzhen & Longgang District People's Hospital of Shenzhen, The Chinese University of Hong Kong, Shenzhen, Guangdong, China.
- Department of Ophthalmology, Baiyun Branch, Nanfang Hospital, Southern Medical University, Guangzhou, China.
- Department of Ophthalmology, Guangdong Eye Institute, Guangdong Provincial People's Hospital and Guangdong Academy of Medical Sciences, Guangzhou, China.
- Department of Ophthalmology, Tohoku University Graduate School of Medicine, Sendai, Japan.
- Department of Ophthalmology, West China Hospital, Sichuan University, Chengdu, China.
ORCID-Profile: Baodi Yang, Hongyang Zhang
Förderung
- Natural Science Foundation of Guangdong Province
- National Natural Science Foundation of China
Zitieren
APA
Yang, B., Lai, C., Qin, Y., Lin, H., Wang, S., Liao, H., & Zhang, H. (2024). Pilot Study on Postoperative Toric Intraocular Lens Alignment. Journal of Ophthalmology, 2024(1). https://doi.org/10.1155/2024/1053914
Vancouver
Yang B, Lai C, Qin Y, Lin H, Wang S, Liao H, et al. Pilot Study on Postoperative Toric Intraocular Lens Alignment. Journal of Ophthalmology. 2024;2024(1). doi:10.1155/2024/1053914
BibTeX
@article{Yang2024,
author = {Yang, Baodi and Lai, Chunxin and Qin, Yongjie and Lin, Hongliang and Wang, Sheng and Liao, Hailan and Zhang, Hongyang},
title = {Pilot Study on Postoperative Toric Intraocular Lens Alignment},
journal = {Journal of Ophthalmology},
year = {2024},
volume = {2024},
number = {1},
doi = {10.1155/2024/1053914}
}
EUVIMED