How Much PEEP Does High Flow Deliver via Tracheostomy? A Literature Review and Benchtop Experiment

Martin Thomas iD, Riddhi Joshi iD, Grant Cave iD

Critical Care Research and Practice · 13.01.2021 · 2021 · 1-9

DOI: 10.1155/2021/6036891 ↗

Kurz gefasst

Übersichtsarbeit in der Zeitschrift Critical Care Research and Practice, erschienen am 13.01.2021. Verfasst von Martin Thomas und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Dubbo Base Hospital, Dubbo, NSW 2830, Australia. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Tamworth Base Hospital.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

High flow tracheostomy (HFT) is a commonly used weaning and humidification strategy for tracheostomised patients, but little is known as to how much PEEP or mechanical benefit it offers. Patient anatomy and device characteristics differentiate it from high flow nasal cannula and the physiological effects observed.

Objectives

(1) To review the available literature on the effects of HFT on airway pressure and indices of gas exchange. (2) To quantify PEEP generated by a HFT circuit.

Methods

A randomised benchtop experiment was conducted, with a size 8 uncuffed Portex tracheostomy connected to an Optiflow™ with Airvo 2™ humidifier system. The tracheostomy tube was partially immersed in water to give rise to a column of water within the inner surface of the tube. An air fluid interface was generated with flows of 40 L/min, 50 L/min, and 60 L/min. The amount of potential PEEP (pPEEP) generated was determined by the distance the water column was pushed downward by the flow delivered.

Findings

Overall 40 L/min, 50 L/min, and 60 L/min provided pPEEP of approximately 0.3 cmH2O, 0.5 cmH2O, and 0.9 cmH2O, respectively. There was a statistically significant change in pPEEP with change in flows from 40–60 L/min with an average change in pPEEP of 0.25–0.35 cmH2O per 10 L/min flow ( p value

Autor:innen und Institutionen

  • Department of Cardiology, Dubbo Base Hospital, Dubbo, NSW 2830, Australia.
  • Department of Intensive Care, Tamworth Base Hospital, North Tamworth, NSW 2340, Australia.
  • Department of Anaesthesia, Dubbo Base Hospital, Dubbo, NSW 2830, Australia.
  • Dubbo Base Hospital, Dubbo, NSW 2830, Australia.
  • Hawkes Bay District Health Board, Hastings, New Zealand.

ORCID-Profile: Martin Thomas, Riddhi Joshi, Grant Cave

Förderung

  • Tamworth Base Hospital

Zitieren

APA

Thomas, M., Joshi, R., & Cave, G. (2021). How Much PEEP Does High Flow Deliver via Tracheostomy? A Literature Review and Benchtop Experiment. Critical Care Research and Practice, 2021, 1-9. https://doi.org/10.1155/2021/6036891

Vancouver

Thomas M, Joshi R, Cave G. How Much PEEP Does High Flow Deliver via Tracheostomy? A Literature Review and Benchtop Experiment. Critical Care Research and Practice. 2021;2021:1-9. doi:10.1155/2021/6036891

BibTeX
@article{Thomas2021,
  author = {Thomas, Martin and Joshi, Riddhi and Cave, Grant},
  title = {How Much PEEP Does High Flow Deliver via Tracheostomy? A Literature Review and Benchtop Experiment},
  journal = {Critical Care Research and Practice},
  year = {2021},
  volume = {2021},
  pages = {1--9},
  doi = {10.1155/2021/6036891}
}
Kostenlos starten →