Evaluation of Stability of Rotating Hinge Knee Prostheses: A Biomechanical Study

Joerg Friesenbichler, Andreas Leithner, Mathias Glehr, Patrick Sadoghi, Werner Maurer-Ertl, Alexander Avian, Reinhard Windhager iD

ISRN Orthopedics · 2013

DOI: 10.1155/2013/701693 ↗

Kurz gefasst

Fachartikel in der Zeitschrift ISRN Orthopedics, 2013. Verfasst von Joerg Friesenbichler und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedic Surgery, Medical University of Graz, Auenbruggerplatz 5, 8036 Graz, Austria. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 3.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Rotating hinge knee prostheses should provide a stable situation following reconstruction. We performed a biomechanical analysis to establish the association between design of the central rotational stem (peg) and implant’s stability, in a theoretical setting.

Methods

Six different rotating hinge designs were tested, and three observers performed two different measurements with a custom made biomechanical apparatus and laterally directed pressure. The aim was to assign the degree of tilting of the peg within the vertical post-in channel by extending the distraction as well as the maximum amount of distraction before the peg’s dislocation. An intraclass-correlation coefficient (ICC) was calculated to determine the observer’s reliability.

Results

Implant designs with cylindrical pegs of different lengths were superior to implant designs with conical or other shaped pegs concerning stability and maximum amount of distraction before dislocation, showing steep rising distraction-angular displacement curves. The ICC at 15 mm and 25 mm of distraction revealed high interobserver reliability ( P < 0.001 ).

Conclusion

The biomechanical analysis showed that rotating hinge prostheses with long and cylindrical pegs have the highest stability at any given amount of distraction. Designs with shorter and markedly tapered pegs may become unstable under conditions of mild joint distraction which has to be proven in future in vivo investigations.

Autor:innen und Institutionen

  • Department of Orthopedic Surgery, Medical University of Graz, Auenbruggerplatz 5, 8036 Graz, Austria.
  • Institute for Medical Informatics, Statistics and Documentation, Medical University of Graz, Auenbruggerplatz 2, 8036 Graz, Austria.
  • Department of Orthopedic Surgery, Medical University of Vienna, Waehringer Gürtel 18-20, 1090 Vienna, Austria.

ORCID-Profile: Reinhard Windhager

Zitieren

APA

Friesenbichler, J., Leithner, A., Glehr, M., Sadoghi, P., Maurer-Ertl, W., Avian, A., & Windhager, R. (2013). Evaluation of Stability of Rotating Hinge Knee Prostheses: A Biomechanical Study. ISRN Orthopedics. https://doi.org/10.1155/2013/701693

Vancouver

Friesenbichler J, Leithner A, Glehr M, Sadoghi P, Maurer-Ertl W, Avian A, et al. Evaluation of Stability of Rotating Hinge Knee Prostheses: A Biomechanical Study. ISRN Orthopedics. 2013. doi:10.1155/2013/701693

BibTeX
@article{Friesenbichler2013,
  author = {Friesenbichler, Joerg and Leithner, Andreas and Glehr, Mathias and Sadoghi, Patrick and Maurer-Ertl, Werner and Avian, Alexander and Windhager, Reinhard},
  title = {Evaluation of Stability of Rotating Hinge Knee Prostheses: A Biomechanical Study},
  journal = {ISRN Orthopedics},
  year = {2013},
  doi = {10.1155/2013/701693}
}
Kostenlos starten →