Design and biomechanical analysis of patient-specific porous tantalum prostheses for knee joint revision surgery
International Journal of Bioprinting · 01.08.2024 · 9 (4) · 735
DOI: 10.18063/ijb.735 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Bioprinting, online erschienen am 01.08.2024. Verfasst von Shilong Mao und 6 weiteren Autor:innen. Beteiligte Einrichtung: College of Bioengineering, Chongqing University; Key Laboratory of Biorheological Science and Technology (Chongqing University), Ministry of Educatio…. Die Arbeit wurde bisher 6-mal zitiert (Crossref). Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Artificial joint revision surgery, as an increasingly common surgery in orthopedics, often requires patient-specific prostheses to repair the bone defect. Porous tantalum is a good candidate due to its excellent abrasion and corrosion resistance and good osteointegration. Combination of 3D printing technology and numerical simulation is a promising strategy to design and prepare patient-specific porous prostheses. However, clinical design cases have rarely been reported, especially from the viewpoint of biomechanical matching with the patient’s weight and motion and specific bone tissue. This work reports a clinical case on the design and mechanical analysis of 3D-printed porous tantalum prostheses for the knee revision of an 84-year-old male patient. Particularly, standard cylinders of 3D-printed porous tantalum with different pore size and wire diameters were first fabricated and their compressive mechanical properties were measured for following numerical simulation. Subsequently, patientspecific finite element models for the knee prosthesis and the tibia were constructed from the patient’s computed tomography data. The maximum von Mises stress and displacement of the prostheses and tibia and the maximum compressive strain of the tibia were numerically simulated under two loading conditions by using finite element analysis software ABAQUS. Finally, by comparing the simulated data to the biomechanical requirements for the prosthesis and the tibia, a patient-specific porous tantalum knee joint prosthesis with a pore diameter of 600 μm and a wire diameter of 900 μm was determined. The Young’s modulus (5719.32 ± 100.61 MPa) and yield strength (172.71 ± 1.67 MPa) of the prosthesis can produce both sufficient mechanical support and biomechanical stimulation to the tibia. This work provides a useful guidance for designing and evaluating a patient-specific porous tantalum prosthesis.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- College of Bioengineering, Chongqing University; Key Laboratory of Biorheological Science and Technology (Chongqing University), Ministry of Education, Chongqing, 400030, China.
- Southwest Hospital of Army Medical University, Chongqing, 400038, China.
- Chongqing Institute of Optics and Mechanics, Chongqing, 401122, China.
Zitieren
APA
Mao, S., Liu, Y., Wang, F., He, P., Wu, X., Ma, X., & Luo, Y. (2024). Design and biomechanical analysis of patient-specific porous tantalum prostheses for knee joint revision surgery. International Journal of Bioprinting, 9(4), 735. https://doi.org/10.18063/ijb.735
Vancouver
Mao S, Liu Y, Wang F, He P, Wu X, Ma X, et al. Design and biomechanical analysis of patient-specific porous tantalum prostheses for knee joint revision surgery. International Journal of Bioprinting. 2024;9(4):735. doi:10.18063/ijb.735
BibTeX
@article{Mao2024,
author = {Mao, Shilong and Liu, Yang and Wang, Fuyou and He, Peng and Wu, Xianzhe and Ma, Xingshuang and Luo, Yanfeng},
title = {Design and biomechanical analysis of patient-specific porous tantalum prostheses for knee joint revision surgery},
journal = {International Journal of Bioprinting},
year = {2024},
volume = {9},
number = {4},
pages = {735},
doi = {10.18063/ijb.735}
}