Dual-Source CT Angiography of Peripheral Arterial Stents: In Vitro Evaluation of 22 Different Stent Types

Michael Köhler, Matthias C Burg, Alexander C Bunck, Walter Heindel, Harald Seifarth, David Maintz

Radiology Research and Practice · 2011

DOI: 10.1155/2011/103873 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiology Research and Practice, 2011. Verfasst von Michael Köhler und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Radiology, University of Münster, Albert-Schweitzer-Stra β e 33, 48129 Münster, Germany. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 3.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To test different peripheral arterial stents using four image reconstruction approaches with respect to lumen visualization, lumen attenuation and image noise in dual-source multidetector row CT (DSCT) in vitro. Methods and Materials . 22 stents (nitinol, steel, cobalt-alloy, tantalum, platinum alloy) were examined in a vessel phantom. All stents were imaged in axial orientation with standard parameters. Image reconstructions were obtained with four different convolution kernels. To evaluate visualization characteristics of the stent, the lumen diameter, intraluminal density and noise were measured.

Results

The mean percentage of the visible stent lumen diameter from the nominal stent diameter was 74.5% ± 5.7 for the medium-sharp kernel, 72.8% ± 6.4 for the medium, 70.8% ± 6.4 for the medium-smooth and 67.6% ± 6.6 for the smooth kernel. Mean values of lumen attenuation were 299.7HU ± 127 (medium-sharp), 273.9HU ± 68 (medium), 270.7HU ± 53 (medium-smooth) and 265.8HU ± 43. Mean image noise was: 54.6 ± 6.3, 20.5 ± 1.7, 16.3 ± 1.7, 14.0 ± 2 respectively.

Conclusion

Visible stent lumen diameter varies depending on stent type and scan parameters. Lumen diameter visibility increases with the sharpness of the reconstruction kernel. Smoother kernels provide more realistic density measurements inside the stent lumen and less image noise.

Autor:innen und Institutionen

  • Department of Clinical Radiology, University of Münster, Albert-Schweitzer-Stra β e 33, 48129 Münster, Germany.

Zitieren

APA

Köhler, M., Burg, M. C., Bunck, A. C., Heindel, W., Seifarth, H., & Maintz, D. (2011). Dual-Source CT Angiography of Peripheral Arterial Stents: In Vitro Evaluation of 22 Different Stent Types. Radiology Research and Practice. https://doi.org/10.1155/2011/103873

Vancouver

Köhler M, Burg MC, Bunck AC, Heindel W, Seifarth H, Maintz D. Dual-Source CT Angiography of Peripheral Arterial Stents: In Vitro Evaluation of 22 Different Stent Types. Radiology Research and Practice. 2011. doi:10.1155/2011/103873

BibTeX
@article{Khler2011,
  author = {Köhler, Michael and Burg, Matthias C and Bunck, Alexander C and Heindel, Walter and Seifarth, Harald and Maintz, David},
  title = {Dual-Source CT Angiography of Peripheral Arterial Stents: In Vitro Evaluation of 22 Different Stent Types},
  journal = {Radiology Research and Practice},
  year = {2011},
  doi = {10.1155/2011/103873}
}
Kostenlos starten →